Büchi Labortechnik AG
Operating the interface pro | 6
Operation Manual UniversalExtractor E-800
57/88
u
Tap the name of the method that you wish to edit.
ð
The display shows the view
Method
.
u
Tap the function
[Edit]
on the function bar.
u
Tap the action
[Sox valve open time]
.
ð
The display shows a dialog with the selectable values.
u
Tap the value you wish to use.
u
Tap the function
[OK]
on the function bar.
ð
The dialog box closes.
u
Tap the function
[Save]
on the function bar.
ð
The Soxhlet valve open time is saved.
6.6.10 Setting the rinse time of a method (Soxhlet, Soxhlet Warm, Hot and
Twisselmann Extraction)
Enter the duration of the rinse step in minutes.
at least 0 minutes
maximum 5940 minutes / 99 hours
Navigation path
➔
u
Navigate to the
[Method]
menu via the navigation path.
u
Tap the name of the method that you wish to edit.
ð
The display shows the view
Method
.
u
Tap the function
[Edit]
on the function bar.
u
Tap the action
[Rinse time]
.
ð
The display shows a dialog box with a numeric input box.
u
Enter the value in the numeric input box.
u
Tap the function
[OK]
on the function bar.
ð
The dialog box closes.
u
Tap the function
[Save]
on the function bar.
ð
The rinse time is saved.
6.6.11 Setting the rinse heating level of a method (Soxhlet, Soxhlet Warm,
Hot and Twisselmann Extraction)
The default setting for the rinse heating level is depending on the selected solvent.
Change the rinse heating level to compensate ambient conditions.
WARNING
Setting the heating level too high
The consequences are device damage, serious injuries or death.
u
Use the pre-set heating levels.
u
Adjust the heating level to compensate the ambient conditions.
Summary of Contents for E-800
Page 1: ...UniversalExtractor E 800 Operation Manual 11593959 C en ...
Page 87: ......