Cleaning and servicing | 8
Operation Manual UniversalExtractor E-800
73/88
Component
Action
Frequency
Heating plate
u
Wipe down the heating plate with a
damp cloth.
u
If heavily soiled, use ethanol or a mild
detergent.
u
If cracked or broken, contact BUCHI
customer service.
Weekly
Hoses and hose con-
nections
u
Check the hoses and hose connections
for defects (cracks, brittle areas).
u
If defective, replace defect hoses.
Weekly
Housing
u
Wipe down the housing with a damp
cloth.
u
If heavily soiled, use ethanol or a mild
detergent.
Weekly
Warning symbols
u
Check that the warning symbols on the
instrument are legible.
u
If they are dirty, clean them.
Weekly
Protection shields
NOTICE! Do not use Acetone
u
Wipe down the protection shields with
a damp cloth.
u
Make sure that they click into place. If
not, replace the gliding elements. See
Replacing the gliding elements
u
If defective or damaged, replace them.
Weekly
Reflector foil for an-
alyte protection
NOTICE! Do not use any detergent
u
Switch the On/Off master switch to Off.
u
Make sure that there are no extraction
beakers in the instrument.
u
Check the reflector foil for dirt.
u
If dirty, wipe down the reflector foil
with a damp cloth.
u
Switch the
On/Off
master switch to On.
Weekly
Interface
u
Wipe down the display with a damp
cloth.
Monthly
8.2
Removing the tank bottle
u
Switch the
On/Off
master switch to Off.
u
Remove the joint clip.