8 | Cleaning and servicing
Büchi Labortechnik AG
74/88
Operation Manual UniversalExtractor E-800
u
Push the solvent tank up.
u
Tilt the solvent tank at 20° and remove it.
8.3
Draining solvent from the instrument
1. Switch on the pump. See Chapter 8.3.1 "Switch on the pump", page 74
2. Open the tank valves. See Chapter 8.3.2 "Open the tank valves", page 74
3. Switch of the valves. See Chapter 8.3.3 "Switch of the pump", page 75
8.3.1
Switch on the pump
Navigation path
➔
➔
[Service]
➔
[Actors]
Precondition:
R
A tank bottle is installed. See Chapter 5.15 "Preparing the tank bottle", page 42
u
Navigate to
[Actors]
via the navigation path.
ð
The display shows the view
Actors
.
u
Enable the action
[Pump]
.
8.3.2
Open the tank valves
Navigation path
➔
➔
[Service]
➔
[Actors]
➔
[Valves]
u
Navigate to
[Valves]
via the navigation path.
ð
The display shows the view
Valves
.
u
Select the position of the tank valve you wish to open.