Büchi Labortechnik AG
Safety | 2
Operation Manual UniversalExtractor E-800
13/88
u
Make sure that the locking mechanism of the protection shield is working
properly.
u
If faults occur:
Switch off the device immediately.
Unplug the power cord.
Inform the operator.
Cool the device before taking any further action.
u
Do not continue to use devices that are damaged.
2.6.4
Glass breakage
Broken glass can cause severe cuts.
Not visible cracks on glassware can lead to breakage.
u
Wear protective gloves when handling broken glassware.
u
Handle the glass components carefully and do not drop them.
u
Always place the glassware in a suitable holder when they are not in use.
u
Prevent any mechanical impact.
u
Prevent thermal shocks on glassware (e.g. do not put hot glassware on cold
surface).
u
Always visually inspect glass components for damage every time they are used.
u
Do not continue to use glass components that are damaged.
u
Do not use beakers which have been heated empty on a heating plate for more
than 20 minutes.
2.6.5
Glass damages during extraction process
Glass damages during the extraction process can cause instrument damages caused
by solvents, coolant and fire.
u
If fault occurs:
Switch off the device immediately.
Unplug the power cord.
Inform the operator.
Let the temperature of the instrument decrease under observation until it is the
same as the ambient.
2.6.6
Overfilled extraction beakers
Overfilled extraction beakers can cause spills and lead to instrument damage or fire.
u
Do not fill more than the indicated maximum quantity. See Chapter 3.5
"Technical data", page 25
2.6.7
Spilled hazardous materials
It is the responsibility of the operator to ensure the safety of users and the selection
of correct decontamination agents.
u
Make sure that the personal protective equipment meets the requirements of
the safety data sheets for the chemicals used.
u
Make sure that the decontamination agent meets the requirements of the safety
data sheets for the chemicals used.
Summary of Contents for E-800
Page 1: ...UniversalExtractor E 800 Operation Manual 11593959 C en ...
Page 87: ......