6 | Operating the interface pro
Büchi Labortechnik AG
56/88
Operation Manual UniversalExtractor E-800
u
Tap the name of the method that you wish to edit.
ð
The display shows the view
Method
.
u
Tap the function
[Edit]
on the function bar.
u
Tap the action
[Chamber heater]
.
ð
The display shows a dialog box with a numeric input box.
u
Enter the value in the numeric input box.
u
Tap the function
[OK]
on the function bar.
ð
The dialog box closes.
u
Tap the function
[Save]
on the function bar.
ð
The drying heating level is saved.
6.6.9
Setting the Soxhlet valve open time of a method (Soxhlet and Soxhlet
Warm Extraction)
The Soxhlet valve open time depends on the position of the level sensor and the
sample structure.
Set the Soxhlet valve opening time so that the extraction glass chamber universal is
completely emptied.
The following open times are available:
Open time
Level sensor
position
Explanation
[short]
bottom
The sample structure is easily releasing
the solvent.
Hydrolysed sample in the glass sample
tubes are extracted.
[mid]
middle
The sample structure is easily releasing
the solvent.
[long]
bottom
The sample structure is slowly releasing
the solvent.
The extraction glass chamber universal
LSV is used.
middle
The extraction glass chamber universal
LSV is used.
top
The sample structure is easily releasing
the solvent.
[max]
top
The extraction glass chamber universal
LSV is used.
The sample structure is slowly releasing
the solvent.
Navigation path
➔
u
Navigate to the
[Method]
menu via the navigation path.