Operation Manual UniversalExtractor E-800
63/88
6.7.7
Changing the molecular weight of a solvent
Navigation path
➔
u
Navigate to the
[Solvent]
menu via the navigation path.
u
Tap the name of the solvent you wish to edit.
ð
The display highlights the selection in green.
u
Tap the function
[OK]
on the function bar.
ð
The display shows the view
Solvent
.
u
Tap the action
[Manually]
.
ð
The display highlights the action
[Mol weight]
in white.
u
Tap the action
[Mol weight]
.
u
Tap the function
[OK]
on the function bar.
ð
The display shows a dialog box with a numeric input box.
u
Enter the mol weight for the solvent.
u
Tap the function
[OK]
on the function bar.
ð
The value for the mol weight is saved.
ð
The dialog box closes.
6.7.8
Changing the density of a solvent
Navigation path
➔
u
Navigate to the
[Solvent]
menu via the navigation path.
u
Tap the name of the solvent you wish to edit.
ð
The display highlights the selection in green.
u
Tap the function
[OK]
on the function bar.
ð
The display shows the view
Solvent
.
u
Tap the action
[Manually]
.
ð
The display highlights the action
[Density]
in white.
u
Tap the action
[Density]
.
u
Tap the function
[OK]
on the function bar.
ð
The display shows a dialog box with a numeric input box.
u
Enter the density for the solvent.
u
Tap the function
[OK]
on the function bar.
ð
The value for the density is saved.
ð
The dialog box closes.