CU-EMA+
5.3.2 Submenu 1
Target temperature (Temperature)
Display → Cooler →
temp
This setting determines the nominal temperature for the cooler temperature.
Parameter range:
2 °C to 20 °C (35.6 °F to 68 °F)
Factory setting:
5 °C (41 °F)
Note:
If the temperature is changed the indicator may blink, until the new operating range has been
reached.
This menu item is hidden if the keylock is enabled.
upper alarm limit (alarm high)
Display → Cooler →
A hi
Here you can set the upper threshold for the visual signal and the alarm relay. The alarm limit is set
based on the cooler temperature setting.
Parameter range:
1 °C to 7 °C (1.8 °F to 12.6 °F)
Factory setting:
3 °C (5.4 °F)
Note:
This menu item is hidden if the keylock is enabled.
lower alarm limit (alarm low)
Display → Cooler →
A Lo
Here you can set the lower threshold for the visual signal and the alarm relay. The alarm limit is set
based on the cooler temperature setting.
Parameter range:
-1 °C to -3 °C (-1.8 °F to -5.4 °F)
Factory setting:
-3 °C (-5.4 °F)
Note:
This menu item is hidden if the keylock is enabled.
Regulator/heated line
Display →
tcMA
→
Adon
This will take you to the target temperature setting for the heated line regulator and the tolerance
range (alarm thresholds).
Enabling/disabling the heated line (Modu)
Display → Cooler →
Modu
Used to de(activate) the heated line.
Parameter range:
nop
,
off
Factory setting:
oFF
Note:
This menu item is hidden if the keylock is enabled.
Exit submenu 1
Display → Submenu →
E
Selecting this will return you to the main menu.