CPsingle, CPdouble
1 Introduction
1.1 Intended use
This unit is intended to discharge condensate from cooled process fluids. The temperature of these mediums is approx. 5 °C.
The unit is suitable for use in normal, non-hazardous areas and according to FM for general areas.
Pump models for the USA and Canada 4492***1*** in non-explosive areas
The peristaltic pumps must be installed inside a housing which requires a tool to open and meets the requirements of the over-
all installation with respect to the housing, layout, space requirement and condensate separation.
Select a housing which meets the requirements of the pump's intended use with respect to mounting, spacing and creepage
paths. The housing must be suitable for operating temperatures of 0 °C to min. 52 °C.
It must be fully wired inside the housing. The cables and terminals used must be US-listed or (if applicable) CSA certified. They
must be designed for the nominal voltage, the nominal current and an operating temperature range of 0 °C to 52 °C.
Water and contaminants must be prevented from entering the unit.
1.2 Scope of delivery
– 1 x Peristaltic pump
– Product documentation
– Connection- and mounting accessories (only optional)
1.3 Type plate
Example:
CPsingle-SA-AC 230/115V 50/60Hz 25/44mA
101001080015 44921120101
Material: PVDF, Norprene
Medium max: see manual; Tamb: 0°C-50°C
Year: 2016 Manual doc.no.: BX450021
Manufacturer and address
Model designation, electrical connection
Purchase order no.+Item no.+Counter
Materials in contact with mediums
Temperature specifications
Year of manufacture, Operating instructions no.
2
Bühler Technologies GmbH
BE450021 ◦ 01/2017