23
Assembly
7.7
Mounting the electronic module to the sensor
holder (without fluid)
→
Make sure that the pH or redox probe is mounted into the sensor holder. Refer to chpt. 7.6.
A
B
C
→
Make sure that seal "A" on the holder is in good condition. Replace it if
necessary. Refer to chpt. 11.
→
Clean connectors "B" and "C" for connection of the pH/redox probe
with alcohol to avoid measurement errors.
→
Insert the electronic module into the holder, making sure the polarising
slots are correctly positioned.
→
Apply slight vertical pressure to engage the seal.
→
Fasten the electronic module and the holder together by tightening the
G2'' nut.
→
Tighten the G2'' nut by hand only, until it stops turning, to ensure good
electrical contact.
Fig. 11: Mounting the electronic module to the holder (without fluid)
→
Mount the display module. Refer to chpt. 7.4.
→
Calibrate the device. Refer to chpt. 9.12.4.
English
Type 8202 ELEMENT