53
Adjustmentandstart-up
9.11.10 setting the parameters of the transistor outputs
See chpt. 9.9 to access the Parameters menu.
Warning
risk of injury due to wrong adjustment.
▶
Before setting the parameters for the display, choose the type of probe mounted on the device. Refer to
chpt. 9.11.11.
This is
when the
device is be-
ing parame-
tered............
....................
Param
This is
when the
device is be-
ing parame-
tered............
....................
TR1 / TR2
Delay:
Contact:
Normally open
Normal. closed
INPUT
Low:
INPUT
High:
INPUT
Outputs
PVar:
pH
TempC
mV_pH
mV_ORP
TempF
Mode:
Hysteresis
Window
warning
If PVar
≠
"warning"
If PVar
≠
"warning"
If PVar
≠
"warning"
PVAR:
choose a process value associated with transistor output 1 or transistor output 2 respectively or associate
the generation of a "warning" event (see chpt. 9.12.4, 9.13.2, 9.13.3 and 9.15.1) to transistor output 1 or tran-
sistor output 2 respectively.
If the selected transistor output is linked to the "warning" event, the transistor switches as soon as such an event
is generated by the device.
See also chpt. "10.5 Solving a problem".
MODE
: choose the operating mode for transistor output 1 or transistor output 2 (see Fig. 38 and Fig. 39).
LOW
: enter the low switching threshold value for transistor output 1 or transistor output 2 (see Fig. 38 and
Fig. 39).
HIGH
Fig. 39).
CONTACT
: choose the type of off-position (normally open, NO, or normally closed, NC) for transistor output 1 or
transistor output 2 (see Fig. 38 and Fig. 39).
DELAY
: choose the value of the time delay prior to switching for each transistor output.
Switching only occurs if one of the thresholds, high or low (functions "High" or "Low"), is exceeded for a duration
longer than this time delay. The time delay before switching is applicable to both output thresholds.
English
Type 8202 ELEMENT