Adjustmentandstart-up
calibrate the redox ("orp") sensor ("calibration" function in the "probe" menu)
• The Redox ("ORP") sensor can be calibrated according to a 1-point procedure.
• In order not to interrupt the process, activate the HOLD function. Refer to chpt. 9.12.1.
• Before each calibration, correctly clean the electrode with a suitable product.
• Set the periodicity of calibrations in the "Interval" function in the sub-menu CALIB INTERVAL (see
heretofore): each time a calibration is due, the device generates a "maintenance" event and a "warning"
event.
1-point calibration
The 1-point calibration procedure is used for quick calibration by adjusting the offset of the measurement graph
with a buffer respectively a calibration solution with a known oxidation reduction potential ("ORP").
English
Type 8202 ELEMENT