9.12.6 Entering an offset for the temperature measurement
See chpt. 9.9 to access the Calibration menu.
The temperature transmitted by the Pt1000 probe may be corrected. This correction value is the temperature
offset.
Calib
Sensor
Temperature
INPUT
Enter the temperature offset
(input limits: ±5 °C)
9.13 Knowing the Diagnostic menu
9.13.1 Modifying the Diagnostic menu access code
See chpt. 9.9 to access the Diagnostic menu.
Code
0***
Confirm code
0***
Diagnostic
System
Enter the new code
Confirm the new code
If the default code (0000) is entered, the code will not be requested to access the menu.
9.13.2 Monitoring the condition of the probe
See chpt. 9.9 to access the Diagnostic menu.
Diagnostic
Glass electrode
Sensor
Impedance
Activate:
Yes/No
READ
Warn lo:
INPUT
Err hi:
INPUT
Err lo:
INPUT
Warn hi:
INPUT
Ref. electrode
Too low or too high an impedance value of one or the other electrode of the measuring probe may indicate that
the probe is broken or used.
Tab. 3 :
Possible combinations of electrode monitoring depending on the measurement mode, symmetrical or
asymmetrical
monitoring of the impedance
ph probe
redox ("orp") probe
glass electrode
reference
electrode
glass electrode
reference electrode
Symmetrical
Possible
Possible
Impossible
Possible
Asymmetrical
Possible
Impossible
Impossible
Impossible