7
AbouttheOperatingInstructions
▶
Indicates an instruction for risk prevention.
→
Indicates a work step that you must carry out.
Indicates a result.
2
iNTENDED usE
use of the device that does not comply with the instructions could present risks to people, nearby
installations and the environment.
The Type 8222 ELEMENT conductivity meter is intended for the measurement of the conductivity in liquids
▶
Use the device in compliance with the characteristics and start-up and use conditions specified in the con-
tractual documents and in the Operating Instructions.
▶
Do not use the device for security applications.
▶
Only operate a device in perfect working order.
▶
Store, transport, install and operate the device properly.
▶
Only use the device as intended.
3
bAsic sAfETy iNforMATioN
This safety information does not take into account any contingencies or occurrences that may arise during instal-
lation, use and maintenance of the device.
The operating company is responsible for the respect of the local safety regulations including staff safety.
risk of injury due to electrical voltage.
▶
Before carrying out work on the system or the device, disconnect the electrical power for all the conductors
and isolate it.
▶
If the device is installed either in a wet environment or outdoors, all the electrical voltages must be of max.
35 V DC.
▶
All equipment connected to the device must be double insulated with respect to the mains according to the
standard UL/EN 61010-1.
▶
Observe all applicable accident protection and safety regulations for electrical equipment.
risk of injury due to pressure in the installation.
▶
Before any intervention in the installation, stop the circulation of fluid, cut off the pressure and drain the pipe.
▶
Before any intervention in the installation, make sure that there is no pressure in the pipe.
▶
Observe the dependency between the fluid temperature and the fluid pressure.
English