burster Italia s.r.l. · Sede legale ed operativa · Via Cesare Battisti, 16/18 · 24035 Curno (BG) · Tel. +39 035 618120 · Fax +39 035 618250
e-mail: info@burster.it · http://www.burster.it
C.F. e P.IVA 11048400151 · Capitale sociale Euro 49.920,00 · R.E.A. di Bergamo 276749
Apparecchi di misura di precisione. Sensori e Sistemi
per grandezze elettriche, termiche e meccaniche
11.1 Proprietary Calibration
Acquisition of measurement points and issue of a calibration protocol Traceable calibrated measuring
equipment is being used for the calibration. The sensor data are being checked during this calibration.
11.2 DKD-Calibration
The calibration of the sensor is carried out according to the guidelines of the DKD and/or DAkkS.. The
surveillance of the calibrating-laboratory takes place by the DAkkS. At this calibration, the uncertainty of
measurement of the torque measuring instrument is determined. Further information can be obtained
from burster.
11.3 Re-Calibration
The recalibration of the torque sensor should be carried out after 26 months at the latest.
Shorter intervals are appropriate:
Overload of the sensor
After repair
After inappropriate handling
Demand of high-quality standards
Special traceability requirements
12