Operation
6.
8
.2
Downloading the configuration
Follow the procedure for downloading
The previous configuration is automatically updated.
All the detector's parameters are downloaded except the following, which must be set by
the operator:
– language
– serial link
– time and date
– temperature unit
– pressure unit.
6.
7
Shutdown the detector
Set the circuit breaker
to
O
(see 4.
3
)
.
Wait for the control panel screen to turn off completely before working on the product,
removing the covers and/or moving it.
Shutdown due to a mains
power failure
When there is a mains power failure, the detector shuts down: it switches on again
au
tomatically when power is restored.
6.
8
Saving and downloading the product’s configuration
When a detector is installed or replaced, it is helpful to copy the configuration (all the pa-
rameters and operating set points programmed by the operator) of a detector that is the
same model.
Do this while the detector is switched on and in Stand-by mode.
6.
8
.1
Saving the configuration
Follow the procedure for saving
(see 7.8.12)
.
The detector's configuration will be saved on the SD card from control panel.
6.
6
Monitoring operation
When the detector is in use, the operator is alerted to incidents as follows:
● Pictogram display indicating that the error message should be read.
● Error display on the screen.
Vacuum the leak detector by pressing the cycle button, then leave it for 2 min in
the cycle before turning it off.