20
Low battery power.
Replace with new batteries and
restart the power.
(See page 7~8)
Incorrect connection of AC
adapter to nebulizer.
Re-connect in the correct
manner and restart the power.
(See page 8)
Power indicator is
lit and nebulizer is
not nebulizing.
Low-power indicator is lit
constantly, insufficient battery
power, or battery has run out.
Replace with new batteries and
restart the power.
(See page 7~8)
Rupture of mesh of medication
chamber.
Replace with a new medication
chamber and then put in the
medication. (See page 15)
Electrodes on medication
chamber are clogged with
medication or water.
Clear the electrodes of clogged
medication or water and restart
the power. (See page 14)
Electrodes on nebulizer and
medication chamber are
stained.
Remove the stains and restart
the power. (See page 14)
Mesh of medication chamber is
severely clogged.
If it still cannot be used after
cleaning, please replace with a
new medication chamber.
(See page 13~15)
Nebulizer shuts off
in usage.
Medication chamber is
loosened and not completely
attached.
Re-attach the medication
chamber correctly and restart
the power. (See page 6)
Connection of AC adapter to
nebulizer is loosened.
Re-connect in the correct
manner and restart the power.
(See page 8)
Medication has run out.
Put in the medication in the
medication chamber.
(See page 9~10)
No contact between
medication and mesh for more
than 10 seconds.
Adjust the nebulizer’s angle so
the medication can come in
contact. (See page 11)
Nebulizer is being shaken in
the use.
Hold the nebulizer in the hand
stably. (See page 11)