GasAlertMicroClip
Options
Low Alarm Acknowledge
Force Bump When Overdue
When
Low Alarm Acknowledge
is enabled, the audible
alarm can be disabled during a low alarm if
C
is pressed, but
the vibrator and alarm LEDs continue to be enabled. This
option is not applicable to the O
2
sensor.
When enabled, the
Force Bump When Overdue
option
forces a bump check if the sensor has exceeded its bump
check interval (see
Bump Interval
). If a successful bump
check has not been performed, the detector will deactivate.
Force Calibration When Overdue
Language Menu
Enabling the
Force Calibration When Overdue
option
forces the detector to enter the calibration process if a sensor
is overdue for calibration upon start-up. If the detector is not
calibrated immediately, the detector will deactivate. The
detector is shipped with this option disabled.
The detector is shipped with English as the default language
displayed. You can choose to view the detector in these
additional languages.
x
French
(
Français
);
x
German
(
Deutsch
);
Cal Lock
x
Spanish
(
Español
); and
If
Cal Lock
is enabled, calibration can only be completed
with an IR device like Soft Tools or the MicroDock II Base
Station. Though
Cal Lock
is enabled, the detector will still
perform an auto zero when activated. The cal lock option is
disabled upon shipment.
x
Portuguese
(
Português
).
21