GasAlertMicroClip
Quick Reference Guide
10
Options
To access the user options, connect the detector to the IR Link
adapter and use the GasAlertMicroClip Soft Tools software.
Refer to the
GasAlertMicroClip Soft Tools manual
for complete
instructions.
The following are the available user options:
1.
Sensors
(H
2
S, CO, O
2
, and LEL)
•
Disabled:
Disables the sensor.
•
Cal Gas:
Changes the span gas concentration for
each sensor.
•
Cal Interval:
Number of days remaining before the
next calibration.
•
Bump Interval:
Number of days remaining before the
next bump check.
•
Low Alarm:
The low alarm setpoint (all sensors).
•
High Alarm:
The high alarm setpoint (all sensors).
•
TWA Alarm:
The time-weighted average (TWA)
alarm setpoint (toxic sensors only).
•
STEL Alarm:
The short-term exposure limit (STEL)
alarm setpoint (toxic sensors only).
•
STEL Interval:
Changes the short-term exposure
limit (5-15 minutes; applicable to toxic sensors only).
•
Auto-Zero on Startup:
Auto zeroes the chosen
sensor during activation (H
2
S, CO, and LEL sensors
only).
•
O2 Auto-Calibration on Startup:
Enables automatic
oxygen calibration upon start-up.
•
LEL By Vol CH4:
Shows the LEL reading in % vol.
assuming a methane environment.
2.
User Options
•
Confidence Beep:
Enables the confidence beep.
•
Latching Alarms:
This option allows an alarm to
remain active until the user acknowledges the alarm.
•
Safe Mode:
Enables the LCD to display
Safe
if the
detector does not enter an alarm.
•
Stealth Mode:
When enabled, the beeper, backlight,
and alarm LEDs are disabled. Only the vibrator
activates.
•
IR Stealth Mode:
When enabled, the beeper and
backlight are disabled. Only the vibrator and IR LEDs
activate.