34
GasAlertQuattro
Quick Reference Guide
Detector does not
accurately measure
gas.
Sensor(s) requires calibration.
Refer to
Calibration on page 10
.
Detector is colder/hotter than gas
temperature.
Allow the detector to attain ambient temperature before use.
Sensor filter is blocked.
Refer to
Replacing the Sensor Filter on page 25
.
Detector does not
enter alarm.
Alarm setpoint(s) are set
incorrectly.
Refer to
Sample Gas Alarm Setpoints
in the
GasAlertQuattro User Manual
.
Define the alarm setpoints in FleetManager II.
Alarm setpoint(s) set to zero.
Refer to
Sample Gas Alarm Setpoints
in the
GasAlertQuattro User Manual
.
Define the alarm setpoints in FleetManager II.
Detector is in calibration mode.
Complete the calibration procedure.
Detector intermittently
enters alarm without
reason.
Ambient gas levels are near
alarm setpoint or the sensor is
exposed to a puff of the target
gas.
Detector is operating normally. Use caution in suspect areas. Check peak
gas exposure reading.
Alarm setpoints are set
incorrectly.
Refer to
Sample Gas Alarm Setpoints
in the
GasAlertQuattro User Manual
.
Define the alarm setpoints in FleetManager II.
Sensor(s) requires calibration.
Refer to
Calibration on page 10
.
Missing or faulty sensor(s).
Refer to
Replacing the Sensors on page 24
.
Features and options
are not operating as
expected.
Changes in FleetManager II.
Verify that the settings in FleetManager II are correct.
Problem
Possible Cause
Solution