background image

2019/08 - Indice de révision : D - Code : 29650

TRAITEMENT DE L'EAU

62/72

9.2 

Electrical problems with the cell

2 types of electrical problems are signaled if detected:

9.2.1  Short circuit in the cell power supply circuit

The message SHRT CELL is displayed at the bottom of the screen.

• 

Cut power to the control panel

• 

Inspect the cable linking the panel to the cell very carefully, starting at the terminal in the  connection 

compartment of the panel. Contact your agent if any damage is visible.

 

!

IMPORTANT

 

This message is also displayed, in parallel with other indicators if the salt concentration 

is too high as described previously. 

9.2.2  No current in the circuit powering the cell 

(eg: the cable is cut, connector not pushed all the way in, connector contact corroded or damaged): the 

message NO CELL is displayed at the bottom of the screen.

• 

Cut power to the cell

• 

Remove the cover over the panel connection terminals, and check that the two lugs on the cell power cable 

are correctly tightened. 

• 

Undo the quick connectors at the cell, check that the contacts inside the connectors are not dirty  (corrosion, 

etc.) 

• 

Check that the connectors are pushed in until a 'click ' is heard.  

• 

Inspect the cable linking the panel and the cell carefully, starting from the terminals in the panel connection 

box. Contact your agent if the cable is damaged.

 

!

IMPORTANT

 This message is also displayed, along with other indicators, if the salt concentration is 

too low as described previously.

9.3 

Flow rate too low

 The message 

NO FLOW

 and the icon 

 appear, indicating that there is either no flow or insufficient flow 

past the flow controller (subject to the reservation that the device is correctly assembled).

 

Stepwise resolution of the problem (if the problem is not resolved after verifying that one step is ok, move 

on to the next step):

• 

Check that the filtration pump motor is running;

• 

Make sure that the filtration pump is primed;

• 

Make sure that suction and return manifold valves in the plant room are not closed;

• 

Make sure that filtration pump prefilter basket is not clogged with impurities; 

• 

Make sure that the filter is not clogged with impurities;

• 

Make sure that the cut-off valves at the bypass loop inlet and outlet are open;

• 

Close the valve between the inlet and the outlet valve on the main circuit completely so that all of the flow 

is directed past the flow sensor;

• 

Cut power to the electrical/ control panel

, remove the lid from the panel terminal connection box, 

disconnect the flow sensor cable, check the condition of the connector contacts, the two wires coming from 

the cable, and the pins on the connector, then push the connector back in fully and the right way around.  

• 

Check the entire length of the flow sensor cable for damage.  

9.4 

Water temperature is too low 

If the water temperature is 

less than 15 °C

, the message 

LOW TEMP

, is displayed:

• 

Between 10 and 15 °C, the maximum Chlorine output possible is 50 %

• 

Below 10°C, Chlorine output cannot rise above 25 %

Summary of Contents for EES

Page 1: ...9 08 Indice de r vision D Code 29650 TRAITEMENT DE L EAU 37 72 INSTALLATION AND OPERATING INSTRUCTIONS to be read carefully and kept for future reference EES ET ESS PRO Electrolysers FR PAGE 1 EN PAGE...

Page 2: ...tion 46 4 5 3 Procedure 47 4 5 4 When starting up for the first time 49 4 6 Installing the sensors 49 4 6 1 The REDOX sensor 49 4 6 2 pH sensor 50 4 7 Installing the electrical control panel 50 4 7 1...

Page 3: ...gh and low ph and redox alarms EES Pro only 63 9 8 1 pH HIGH 63 9 8 2 PH LOW 64 9 8 3 ORP HIGH 64 9 8 4 ORP LOW 64 9 9 Ph or redox sensor faults EES Pro only 64 9 10 pH Redox circuit board fault EES P...

Page 4: ...n which it is mounted These parameters will need to be measured regularly and the pH will need to be adjusted when necessary With the acid dosing pump option installed EES can inject acid to lower the...

Page 5: ...t When handling acid and during interventions on the pump or its tubing protective glasses and clothing should be worn to protect against splashes In the event of contact with the skin flush the area...

Page 6: ...EES Pro Electrical control panel 1 The panels of all models have the same outer appearance and the same dimensions Nevertheless the electrical spcifications functional specifications and the internal...

Page 7: ...nject acid into the hydraulic circuit The acid dosing pump kit includes 2 saddles 50 63mm 3 8 to 1 2 adaptor a check valve valve internal replacement tube 2 x 3m polyethylene tube foot filter and weig...

Page 8: ...remains in the cell In some configurations this will require installation of a check valve The flow sensor should be positioned such that the acid used to descale the cell will not touch it A check v...

Page 9: ...d air Pr cautions vis vis des sondes Les parties immerg es des sondes doivent toujours rester immerg es eau de piscine ou du robinet m me hors fonction ne doivent jamais tre touch es avec les doigts n...

Page 10: ...acid pump and saddle clamp 2 Chlorinator Cell 3 Tee with pH and ORP sensors 4 Flow switch Tee 4 5 2 Horizontal installation 1 Dosing acid pump and saddle clamp 2 Chlorinator Cell 3 Tee with pH and OR...

Page 11: ...re you wish to install the acid pump on the wall Check that the all components electrical wiring tubes etc reach their position Drill and attach the acid pump s clip to the wall and hang the pump on i...

Page 12: ...mm wrench Wrap the threads of the saddle clamp adapter to with Teflon tape and insert it into the saddle clamp Wrap the check valve threading with Teflon tape andinsert it into the adapter Cut the tub...

Page 13: ...ump in MANUAL mode for a few minutes Make sure the red light on the pump is energised that it is running and that acid is being sucked from the acid container and expelled through the injection tube 4...

Page 14: ...end of the sensor into the union until the end of the sensor is located in the centre of the pipe through which the water is flowing Screw the nut firmly onto the union and check that the sensor canno...

Page 15: ...5 m tres mais dans tous les cas plus de 2 m tres du plan d eau sous r serve que son c ble d alimentation soit directement reli un dispositif di rentiel r siduel d clenchant 30 mA exigences norme NF C...

Page 16: ...ory with acid dosing LN2 Live slaved to filtration Cover panel dry contact EES output is unchanged is no voltage is detected at the contact Flow switch Water T C sensor Cell pole Cell pole N Dosing pu...

Page 17: ...panel and acid dosing pump LD2 slaved live cell Nota bene to avoid electrical disturbances when the pump starts and stops the EES power cables should not be in contact with those of the filtration pu...

Page 18: ...buffer tank if the pool has an overflow feature and the initial salt concentration measured beforehand USE THE FOLLOWING TABLE 0 500 1000 1500 2000 2500 3000 3500 4500 10 40 35 30 25 20 15 10 5 0 20 8...

Page 19: ...e message OFF is displayed on the LCD screen Press the ON OFF button 7 If the flow of water past the flow sensor is adequate 5 m3 h the message ON EES or AUTO EES Pro indicating that production of Chl...

Page 20: ...he EES models only the chlorine output can be adjusted manually by the user by pressing the two arrow keys and This setting will be saved to memory for when the device is started in the future even if...

Page 21: ...nd maintain it at this temperature if the pool is not equipped with a heating system simply wait until the water naturally reaches the lowest temperature at which bathers will use the pool Next adjust...

Page 22: ...to switch to Turbo mode the word Turbo blinks at the bottom of the output scale The Turbo mode duration is set to 12 hours in the factory this setting is displayed on screen Successive presses on the...

Page 23: ...ion of Chlorine is decreased to 20 by default The value by which Chlorine production is decreased while the pool is covered may be modified by the user using the arrow keys and The new value will be s...

Page 24: ...n EES the cell must be removed from the hydraulic circuit for descaling 8 7 Punctual or periodic reduction of the ph by the acid dosing pump This function concerns EES Pro and also EES if the acid dos...

Page 25: ...f salt In the event of a low salt warning check that the electrodes are clean and check the salt concentration with a suitable tester in case of doubt A faulty temperature sensor could also cause a lo...

Page 26: ...along with other indicators if the salt concentration is too low as described previously 9 3 Flow rate too low The message NO FLOW and the icon appear indicating that there is either no flow or insuff...

Page 27: ...isplayed at the bottom of the LCD screen Cut power to the electrical control panel open the lid of the connection terminal compartment check the condition of the connector contacts the two wires comin...

Page 28: ...for the Chlorine concentration to fall and to see if the phenomenon reoccurs 9 8 4 ORP LOW ORP LOW this message is displayed when the value measured by the Redox sensor is less than 525 mV If no previ...

Page 29: ...agitate them Rinse the sensors thoroughly after degreasing After being descaled and degreased the sensors should be recalibrated see the next paragraph and the correspondence between the Redox potent...

Page 30: ...the filtration pump off ST BY mode close the valves of the by pass loop on which the tee housing the sensors is mounted place a bucket under the tee to collect any water Remove the pH and the Redox s...

Page 31: ...the tube until it is completely out Note It is advisable to flush the pump with running tap water in order to dilute any remaining acid which may have spilled from the tube Disconnect both ends of the...

Page 32: ...or 10 Hydrochloric acid into the cell Wait until the scale is completely dissolved stops frothing If scale is still present empty the cell and repeat the procedure with vinegar or acid Rinse the cell...

Page 33: ...erizing chemicals are added to the water to prevent the growth of algae andfiltration is run every day Filtration must notably be run while the ambient temperature in the plant room where the material...

Page 34: ...is less than 10 h day The salt concentration is barely within range Measure the salt concentration and if necessary adjust it so that it is between 4 5 and 5 0 g l Scale is starting to build up on th...

Page 35: ...l remove the lid covering the panel connection terminals and tighten the power cable connections The fuse protecting the main circuit board has blown Contact the after sales service for inspection and...

Page 36: ...2019 08 Indice de r vision D Code 29650 TRAITEMENT DE L EAU 72 72 S A au capital de 7 000 000 R C S Rennes B 333263846000 37...

Reviews: