background image

2019/08 - Indice de révision : D - Code : 29650

TRAITEMENT DE L'EAU

64/72

9.8.2  PH LOW 

This message is displayed when the pH measured by the pH sensor is

 less than 6.7.

 

If no previous event explains this situation, try to resolve the problem by carrying out the following steps 

in the order indicated (if the problem is not resolved by one step, proceed with the next):

• 

Make sure that the pH set point is correct.  

• 

Check that the acid dosing pump stops injecting acid when the pH falls below the set point.

• 

Check that the sensor is clean and undamaged.

• 

Verify the accuracy of the pH sensor by measuring the pH of the pool water using a recently calibrated 

photometer that measures pH. If there is a difference of more than 0.2 pH units, calibrate the pH sensor 

and check again.

• 

If the values measured by EES and the photometer correspond, and the pH is less than 6.7, add base to 

bring the pH back to the set point and wait to see if this phenomenon reoccurs. 

9.8.3  ORP HIGH 

This message is displayed if the value measured by the Redox sensor 

greater than 

925 mV

 .

If no previous event explains this situation, try to resolve the problem by carrying out the following steps 

in the order indicated (if the problem is not resolved by one step, proceed with the next):

• 

Make sure that the Redox set point value is correct.

• 

Check that the device is not in Turbo mode and that it is not producing Chlorine

• 

Using a recently calibrated photometer, measure the concentration of free Chlorine in the pool water  to see 

if this value corresponds with the (abnormally high) Redox potential displayed.

• 

Calibrate the Redox sensor

• 

Wait for the Chlorine concentration to fall and to see if the phenomenon reoccurs.

9.8.4  ORP LOW

ORP LOW:

 this message is displayed when the value measured by the Redox sensor is less than 525 mV.

If no previous event explains this situation, try to resolve the problem by carrying out the following steps 

in the order indicated (if the problem is not resolved by one step, proceed with the next):

• 

Make sure that the Redox set point value is correct.

• 

Check that the sensor is clean and undamaged.

• 

Using a recently calibrated photometer, measure the concentration of free Chlorine in the pool water  to see 

if this value corresponds with the (abnormally low) Redox potential displayed.

• 

Calibrate the Redox sensor

• 

Check that filtration is running for long enough each day (at least 10 hours a day during the pool season)

• 

Visually inspect the cell to verify that it is producing and that the electrodes are not covered in scale. 

• 

Check the stabiliser (Isocyanuric acid) concentration in the pool. It should be between  30 and 50 mg/l 

• 

Increase the concentration of free Chlorine and wait to see if this phenomenon reoccurs.  

All these alarms are automatically reset when the parameter returns to within an acceptable range.  

9.9 

pH or Redox sensor faults (EES Pro only)

PH Probe

: pH sensor not detected. 

ORP Probe

: Redox sensor not detected.

• 

Cut power to the electrical/ control panel

, open the connection terminal compartment and check that the 

two sensors are properly connected at the correct positions on the PCB.  

• 

Remove the sensor PCB from its housing and check that the pins are clean and in good condition. Reinsert 

it fully into its housing.

• 

Inspect the entire length of the wires of each sensor and make sure that they are not damaged. 

• 

Close the by-pass loop shut-off valves and remove the sensors to check them for damage.

Summary of Contents for EES

Page 1: ...9 08 Indice de r vision D Code 29650 TRAITEMENT DE L EAU 37 72 INSTALLATION AND OPERATING INSTRUCTIONS to be read carefully and kept for future reference EES ET ESS PRO Electrolysers FR PAGE 1 EN PAGE...

Page 2: ...tion 46 4 5 3 Procedure 47 4 5 4 When starting up for the first time 49 4 6 Installing the sensors 49 4 6 1 The REDOX sensor 49 4 6 2 pH sensor 50 4 7 Installing the electrical control panel 50 4 7 1...

Page 3: ...gh and low ph and redox alarms EES Pro only 63 9 8 1 pH HIGH 63 9 8 2 PH LOW 64 9 8 3 ORP HIGH 64 9 8 4 ORP LOW 64 9 9 Ph or redox sensor faults EES Pro only 64 9 10 pH Redox circuit board fault EES P...

Page 4: ...n which it is mounted These parameters will need to be measured regularly and the pH will need to be adjusted when necessary With the acid dosing pump option installed EES can inject acid to lower the...

Page 5: ...t When handling acid and during interventions on the pump or its tubing protective glasses and clothing should be worn to protect against splashes In the event of contact with the skin flush the area...

Page 6: ...EES Pro Electrical control panel 1 The panels of all models have the same outer appearance and the same dimensions Nevertheless the electrical spcifications functional specifications and the internal...

Page 7: ...nject acid into the hydraulic circuit The acid dosing pump kit includes 2 saddles 50 63mm 3 8 to 1 2 adaptor a check valve valve internal replacement tube 2 x 3m polyethylene tube foot filter and weig...

Page 8: ...remains in the cell In some configurations this will require installation of a check valve The flow sensor should be positioned such that the acid used to descale the cell will not touch it A check v...

Page 9: ...d air Pr cautions vis vis des sondes Les parties immerg es des sondes doivent toujours rester immerg es eau de piscine ou du robinet m me hors fonction ne doivent jamais tre touch es avec les doigts n...

Page 10: ...acid pump and saddle clamp 2 Chlorinator Cell 3 Tee with pH and ORP sensors 4 Flow switch Tee 4 5 2 Horizontal installation 1 Dosing acid pump and saddle clamp 2 Chlorinator Cell 3 Tee with pH and OR...

Page 11: ...re you wish to install the acid pump on the wall Check that the all components electrical wiring tubes etc reach their position Drill and attach the acid pump s clip to the wall and hang the pump on i...

Page 12: ...mm wrench Wrap the threads of the saddle clamp adapter to with Teflon tape and insert it into the saddle clamp Wrap the check valve threading with Teflon tape andinsert it into the adapter Cut the tub...

Page 13: ...ump in MANUAL mode for a few minutes Make sure the red light on the pump is energised that it is running and that acid is being sucked from the acid container and expelled through the injection tube 4...

Page 14: ...end of the sensor into the union until the end of the sensor is located in the centre of the pipe through which the water is flowing Screw the nut firmly onto the union and check that the sensor canno...

Page 15: ...5 m tres mais dans tous les cas plus de 2 m tres du plan d eau sous r serve que son c ble d alimentation soit directement reli un dispositif di rentiel r siduel d clenchant 30 mA exigences norme NF C...

Page 16: ...ory with acid dosing LN2 Live slaved to filtration Cover panel dry contact EES output is unchanged is no voltage is detected at the contact Flow switch Water T C sensor Cell pole Cell pole N Dosing pu...

Page 17: ...panel and acid dosing pump LD2 slaved live cell Nota bene to avoid electrical disturbances when the pump starts and stops the EES power cables should not be in contact with those of the filtration pu...

Page 18: ...buffer tank if the pool has an overflow feature and the initial salt concentration measured beforehand USE THE FOLLOWING TABLE 0 500 1000 1500 2000 2500 3000 3500 4500 10 40 35 30 25 20 15 10 5 0 20 8...

Page 19: ...e message OFF is displayed on the LCD screen Press the ON OFF button 7 If the flow of water past the flow sensor is adequate 5 m3 h the message ON EES or AUTO EES Pro indicating that production of Chl...

Page 20: ...he EES models only the chlorine output can be adjusted manually by the user by pressing the two arrow keys and This setting will be saved to memory for when the device is started in the future even if...

Page 21: ...nd maintain it at this temperature if the pool is not equipped with a heating system simply wait until the water naturally reaches the lowest temperature at which bathers will use the pool Next adjust...

Page 22: ...to switch to Turbo mode the word Turbo blinks at the bottom of the output scale The Turbo mode duration is set to 12 hours in the factory this setting is displayed on screen Successive presses on the...

Page 23: ...ion of Chlorine is decreased to 20 by default The value by which Chlorine production is decreased while the pool is covered may be modified by the user using the arrow keys and The new value will be s...

Page 24: ...n EES the cell must be removed from the hydraulic circuit for descaling 8 7 Punctual or periodic reduction of the ph by the acid dosing pump This function concerns EES Pro and also EES if the acid dos...

Page 25: ...f salt In the event of a low salt warning check that the electrodes are clean and check the salt concentration with a suitable tester in case of doubt A faulty temperature sensor could also cause a lo...

Page 26: ...along with other indicators if the salt concentration is too low as described previously 9 3 Flow rate too low The message NO FLOW and the icon appear indicating that there is either no flow or insuff...

Page 27: ...isplayed at the bottom of the LCD screen Cut power to the electrical control panel open the lid of the connection terminal compartment check the condition of the connector contacts the two wires comin...

Page 28: ...for the Chlorine concentration to fall and to see if the phenomenon reoccurs 9 8 4 ORP LOW ORP LOW this message is displayed when the value measured by the Redox sensor is less than 525 mV If no previ...

Page 29: ...agitate them Rinse the sensors thoroughly after degreasing After being descaled and degreased the sensors should be recalibrated see the next paragraph and the correspondence between the Redox potent...

Page 30: ...the filtration pump off ST BY mode close the valves of the by pass loop on which the tee housing the sensors is mounted place a bucket under the tee to collect any water Remove the pH and the Redox s...

Page 31: ...the tube until it is completely out Note It is advisable to flush the pump with running tap water in order to dilute any remaining acid which may have spilled from the tube Disconnect both ends of the...

Page 32: ...or 10 Hydrochloric acid into the cell Wait until the scale is completely dissolved stops frothing If scale is still present empty the cell and repeat the procedure with vinegar or acid Rinse the cell...

Page 33: ...erizing chemicals are added to the water to prevent the growth of algae andfiltration is run every day Filtration must notably be run while the ambient temperature in the plant room where the material...

Page 34: ...is less than 10 h day The salt concentration is barely within range Measure the salt concentration and if necessary adjust it so that it is between 4 5 and 5 0 g l Scale is starting to build up on th...

Page 35: ...l remove the lid covering the panel connection terminals and tighten the power cable connections The fuse protecting the main circuit board has blown Contact the after sales service for inspection and...

Page 36: ...2019 08 Indice de r vision D Code 29650 TRAITEMENT DE L EAU 72 72 S A au capital de 7 000 000 R C S Rennes B 333263846000 37...

Reviews: