- 4 -
contacted with the specimen. Turn on the testing machine to make the piston rising slowly. Press
“Clear” key to remove the tare.
②
.
Press “Func” key until the indicating lamp test flashed and press “Enter” key. “P 0 0 0 1” “S — —
— 2” will be displayed on the windows.
Where: “P0001” is the number of the specimen and this number can be changed by entering
the number required, and “S---2” is the code for the cross-section of the specimen
1- for 100 mm cubic compression specimen
2- for 150 mm cubic compression specimen (standard specimen)
3- for 200 mm cubic compression specimen
4- for non-standard compression specimen
5- for 150x150x550 mm flexural specimen
6- for 100x100x400 mm flexural specimen
7- 40x40x160 mm flexural specimen
8- for70.7 mm cubic compression specimen
③
.
If “0128” is input, “P0128” will be displayed on the Force Value window. Press “Enter” key for
cross-section of the specimen setting.
④
.
input the code of the specimen. e.g. if the 150 mm cubic specimen is placed on, input “2” and
then press “Enter” key to start a trial test. If the non-standard specimen is placed on, input “4”
and then press “Enter” key to start a trial test. “000.0” will be displayed on the Loading Rate
window. The cross-section of the specimen can be input directly (unit: cm2). If the cross-section
of the specimen is 123.5 cm2, input “1235” and then press “Enter” key to start a trial test.
⑤
.
First, place on the specimen, and then turn off the oil returning valve and turn on the oil feeding
valve to apply on the specimen in a fixed rate based on the testing regulations until the specimen
is crashed. Turn off the oil feeding valve and turn on the oil returning valve at the same time to
relief the load. Press “0” to delete the data tested if the data is not correct.
⑥
.
Place the second specimen on the testing machine after the Force Value window is zeroed.