Instructions for using HoloLens with VimedixAR™
21
©2017 CAE
905K800022 v4
Calibration Procedure
VimedixAR supports the following two methods to calibrate the position of the Hologram Manikin relative to the
Physical Manikin:
•
calibration by tag
•
calibration by gaze
Note
: Calibration by tag or gaze can always be repeated if the Physical Manikin is moved.
Regardless of your Shared Experience Mode and whether you are calibrating by tag or gaze, you have access to the
Information
,
Calibration
and
Help
tabs as well as the
Adjust Menu
button at any point during calibration; all
other tabs and
Hologram Controls
are disabled.
Calibration by Tag
Calibration by tag is started by selecting the
Use Tag
button at the bottom of the
Interactive Panel
after selecting
the
Calibration
tab.
Once the
Use Tag
button is selected, a small target sign appears at the top left corner of your field of view with the
text "Place the hologram using the tag". You must then place the provided "T" tag on the table, below the right ear
of the Physical Manikin with the arrow facing the head. When the HoloLens recognizes the tag, it places a
holographic image of the tag on top of it, with the Hologram Manikin not far. You must then move the tag to change
the position of the Hologram Manikin to match the Physical Manikin. Please refer to the Fine-Tuning Your
Calibration subsection of the user guide to learn how to fine-tune your calibration.
Select the
End
button at the bottom of the
Interactive Panel
to end the calibration by tag.
Calibration by Gaze
Calibration by gaze is started by selecting the
Use Gaze
button at the bottom of the
Interactive Panel
after
selecting the
Calibration
tab.
Once the
Use Gaze
button is selected, a small target sign appears at the top left corner of your field of view with the
text "Air Tap to place the hologram". You must then move the Hologram Manikin with your gaze to match the
position of the Physical Manikin and air tap the Hologram Manikin to anchor it in place. Please refer to the Fine-
Tuning Your Calibration subsection of the user guide to learn how to fine-tune your calibration.
Select the End button at the bottom of the Interactive Panel to end the calibration by gaze.