NGPS – User’s Manual
7
Safety information
The following table shows the general environmental requirements for a correct
operation of instruments referred in this User’s Manual:
The following symbols are used within this manual or are reported on the NGPS
box and along this manual:
CAUTION Risk of Electrical Shock
Caution: Documentation must be consulted in all cases
where this symbol is marked
Indicates ground terminal
Protective Ground Conductor Terminal
Environmental Conditions
Requirements
Environment
Indoor use
Operating Temperature
0°C to 40°C
Operating Humidity
20% to 80% RH (non-condensing)
Altitude
Up to 2000 m
Pollution degree
2
Overvoltage Category
II
Storage Temperature
-10°C to 60°C
Storage Humidity
5% to 90% RH (non-condensing)
Water Cooled Model
Max Water Input Pressure
(continuous)
8 bar
Water Pressure drop
1.5 bar
Water Flow
1.5 l/min