Section 2: Getting Started
A Secure Digital (SD) card and a battery must be installed to operate the VX3+. The SD card
must be installed before the battery is installed.
Secure Digital (SD) Card
The VX3+ stores acquired ECG data to a standard SD card with capacity between 128mb and
2GB. Industrial style SD cards are suggested. To insert the SD card, remove the battery cover
on the back of the recorder and insert the card face up as shown in Fig 2.1. Push it into the slot
until it clicks into place. To remove the SD card, push the card into the slot and it will release.
You can then pull the card out of the slot. The battery must be removed prior to inserting or
removing the SD card.
Battery
The VX3+ uses one alkaline AA battery (IEC-LR6). No other batteries may be used. To install
the battery, remove the battery door on the back of the device, and place as shown on the
placement reference molded on the inside of the battery compartment.
If a battery has power, but is too weak to run a 24-hour study, the recorder will beep continuously
and display a low battery warning. The VX3+ monitors battery voltages during a study. It will run
until the charge falls below an acceptable level. When this occurs the VX3+ will shut down.
Note: A loss of up to 2 minutes data may occur if the battery is removed during a recording.