Personal Ambulatory Oxygen System
Provider Technical Manual
43
REMOVE AND REPLACE 9 VOLT BATTERY
Replace the 9-volt battery when the unit beeps three times at the end of power-on self-test, when voltage is less than 7.0
Volts, or during annual PM. To replace the 9-volt battery, follow these steps:
1. Remove the Eclipse front cover assembly.
2. Lift the battery out of the compartment located at the bottom right corner of the unit.
Figure 17: Removing the 9-volt Battery
3. Disconnect the 9-volt battery harness.
4. Remove and replace the 9-volt battery.
Figure 18: Disconnecting the 9-volt Battery.
5. Reconnect the battery harness. Ensure that the terminals are aligned correctly.
6. Place the battery back into the compartment.
7. Install battery foam pad if necessary.
Disconnect power supplies and remove Power Cartridge before
removing the unit cover. Do not touch exposed circuits during
maintenance without ESD protection.
WARNING:
Battery
Compartment