To replace the compressor intake filter, follow these steps:
1. Cut the cable tie on the silicone tubes attached to the compressor intake filter.
Figure 27: Removing the Compressor Intake Filter.
2. Remove the silicone tubes attached to each end of the filter body.
3. Install a new filter by pushing each tube completely over the barb on the filter body. Verify
flow direction is correct.
4. Reconnect silicone tubes to the filter. Ensure that the feed tubes are not blocked, crimped
or kinked upon completion of the installation or the unit will alarm for low oxygen purity
after the warm-up cycle is complete.
Figure 28: Installing the Compressor Intake Filter.
5. Re-install two cable ties on silicone tubes.