Personal Ambulatory Oxygen System
Provider Technical Manual
58
RED LIGHT
Symptom
Probable Cause
Remedy
During Operation:
Red Light “Blink-
ing” and “3 Beeps”
(Low O2 condition
- Less than 70%)
Press NO SMOKING
button for Diagnostic
Code “ALRM=004”
Restricted Air Flow through cabinet
inlet filter
Perform Annual Preventive Maintenance (5022-SEQ), if issue remains call
Chart TS for service.
Blocked air flow to cabinet inlet
filter
Ensure the Eclipse has adequate air flow around the device (6" around the
device) - turn the device around in the car(exhaust facing to the middle of
the car)
Restricted air flow through Internal
Filters
Perform Annual Preventive Maintenance (5022-SEQ), if issue remains call
Chart TS for service.
Possible Blockage or Leak of air
flow through internal components
Contact Chart TS for RMA service.
During Operation:
Red Light
“Constant” and alarms
“constant tone”
FAIL Condition
Displayed “FAIL XX”
Unit has shut down and displays a
FAIL XX code
May indicate a Power Loss Condition or Internal malfunction.
*Remove all power (Ext power and Battery cartridge) from the Eclipse.
*Wait 15 Seconds.
*Re-Connect External AC power and listen for the External Power connec-
tion "3 second beep". Press the "ON" button and allow the Eclipse to come
up to Green Light operation.
Other
Contact Chart TS for RMA service
Eclipse Red Light
and Alarms:
Battery
has become Too Hot.
FAIL Condition
Displayed "FAIL XX"
Battery is too Hot and will not
operate the Eclipse; above 158°F
(70°C)
*Remove the battery and allow to cool
*Replace with a fresh cool battery.
Eclipse Overheated, Display shows
a FAIL 92, Fail 91, or FAIL A4
Allow the Eclipse to cool off (30 Minutes Minimum) below the internal
temperature of 115 F - remove power, wait 15 seconds, plug back in to
External Power only and restart the Eclipse. Install battery if needed.
Other
Contact Chart TS for RMA service.
BLINKING & BEEPING
STEAD
Y &
CONST
ANT AL
ARM
STEAD
Y &
CONST
ANT AL
ARM
OTHER CONDITIONS
Symptom
Probable Cause
Remedy
Pulse Mode:
Green
Pulse Mode Led is
Steady then Flashing
Press NO SMOKING
button for Diagnos-
tic Code “ALRM=200”
(Eclipse is going from
Pulse to Continuous
every 15 seconds.)
Not detecting a patient's breath
Unit is in a Breath Search mode
Ensure Cannula Tubing is not more than 7ft (2.1m) in length. Ensure it is not
kinked or disconnected.
Ensure you are not using a Humidifier - check for water block in cannula
Ensure Patient is breathing through their nostrils. Patient may be a mouth
breather, not allowing enough negative pressure to trigger the device's
bolus volume.
Provider to properly set the Pulse Sensitivity settings for their client. (menu:
PS=XX)
Other
Contact Chart TS for RMA service