Personal Ambulatory Oxygen System
Provider Technical Manual
59
OTHER CONDITIONS (CONTINUED)
Battery Power:
My
Battery is not Lasting
the time it used to…
NOTE: Battery times
will decrease with
higher bolus size,
breath rate, ambient
temperature, Battery
age and use over time.
Battery End of Life: May have exceeded
the life expectancy of 500 full Charge/
Discharge cycles.
Discharge it fully, recharge it fully, then measure the length of discharge to
within 20% of new specification. Replace if higher that 20% loss of charge.
Calibration: The Eclipse Battery may be
out of calibration
*Unplug the Eclipse from external
*Allow Eclipse to operate until the Red Light comes on and alarm sounds
constant(Loss or Power)
*Plug in to External AC power and allow to recharge
The Eclipse display may display (****)
stars in place of the battery display
Indicates a loss of communication from the battery. Change the Battery
with a fresh one. If the issue remains, contact your home oxygen provider
for service.
A power interruption that causes The
Eclipse display to read "FAIL 81", or
"FAIL 95"
*Red Light and constant tone alert
This error code may indicate a Power Loss and reconnection. *Remove all
power(Ext power and Battery cartridge) from the Eclipse. *Wait 15 Seconds.
*Re-Connect External AC power and listen for the External Power connec-
tion "3 second beep". Press the "ON" button and allow the Eclipse to come
up to Green Light operation.
Other
Contact Chart TS for RMA service
Eclipse runs on Battery
power when plugged
into external power
(AC or DC)
Unit will beep for loss
of external power
AC power is loose or not connected
Plug device directly into known good AC outlet, extension cord and com-
puter strips are not recommended
DC power is loose or not connected
*Remove battery from the eclipse and ensure can operate the Eclipse un-
interrupted on car power
*Use power cartridge(s) if car does not have adequate power
*Have outlet checked for proper current rating
Unit runs the battery down when you
don’t expect it to?
*Unit may be plugged into external power but the power verification light
is "not" on
*Check outlet, check plug on power supply, check power plug on Eclipse
Other
Contact Chart TS for RMA service
Power Cartridge
(Battery) status gauge
no longer indicates
fully charged
Power Cartridge (battery) is partially/
completely discharged.
Plug device directly into known good AC outlet
(verify a waterfalling
[scrolling bottom to top] battery indicator)
Power Cartridge (battery) is not seated
properly in the Eclipse compartment.
*Reseat Power Cartridge in the Eclipse battery compartment
*Ensure both locking pins are engaged [check for waterfalling (scrolling
bottom to top) battery indicator if not fully charged]
Power Cartridge (battery) out of
Calibration.
Calibrate Battery:
Unplug ext power turn unit on, allow to run until
power loss (Red Light and constant tone alert), plug unit into AC power to
charge (ensure indicator is waterfalling [Scrolling Bottom to top]).
Power Cartridge (battery) is aging.
Typical average end of life is 500 charge/discharge cycles:
Replace
with a known good battery
Other
Contact Chart TS for RMA service.
VISU
AL
VISU
AL
ONE BEEP ONL
Y
WARNING: Do not modify this equipment without authorization of the
manufacturer.