Personal Ambulatory Oxygen System
Provider Technical Manual
7
Pulse Dose Mode Specifications
Pulse Settings
1.0 to 6.0, in 16mL increments and 128mL, 160mL, 192mL
Trigger Sensitivity
Adjustable between settings of 1 (most sensitive) to 3 (least sensitive)
Adjustable Bolus Rise Time
Adjustable settings of Fast(factory setting), Medium, or Slow
Trigger Criteria
• Cannula pressure has dropped below the trigger point (typically between 0.15– 0.45
cm of H
2
O of negative pressure)
• At least 1¼ seconds has passed since the last pulse began
Minimum time between
breaths
1.25 seconds (max. 3 consecutive breaths)
Response to Missing Breaths
While in Pulse Dose Mode, the Eclipse 3 is always monitoring for breath detection.
After 15 seconds of no breath detected, the Eclipse 3 “delivers” Continuous Flow at the
last Continuous Flow setting. After another 15 seconds, the Eclipse 3 stops delivering
Continuous Flow and waits 15 seconds for a breath. The Eclipse 3 will stay in this
modality until a breath is detected. (See Power Cartridge Conservation Feature, page 26.)
Power Accessory Specifications
AC Power Supply
Input Voltage
100-240VAC, 50-60 Hz
Input Power
245-260 VA
Output Voltage
28 VDC
Output Power
200W
DC Power Supply
Input Voltage
11.5-18VDC
Output Voltage
26 VDC
Output Power
150W Max
Power Cartridge (Battery)
Output voltage
14.8 VDC
Capacity
Quantity (2) 97.5 W-hrs batteries
(Each containing 7.92 grams equivalent Lithium content)
Nominal Power Cartridge Life
80% Capacity after 500 Charge/Discharge cycles
Power Cartridge Recharge Time
1.8 to 5.0 hours, dependent on flow setting, to achieve 80% capacity
from a fully discharged Power Cartridge
NOTE: Bolus volume decreases as breath rate exceeds published range.
B
O
LU
S V
O
LU
M
E
(+
/-
15
%)
Pulse Dose
Setting
Bolus Size
(± 15%) mL
AC Power Supply and
Power Cartridge (Battery)
DC Power Supply
Max Breath
Rate
Max Breath
Rate
1.0
16
40
40
2.0
32
40
40
3.0
48
40
40
4.0
64
40
31
5.0
80
37
25
6.0
96
31
20
-
128
23
15
-
160
18
12
-
192
15
10