Personal Ambulatory Oxygen System
Provider Technical Manual
80
REMOVE AND REPLACE THE POWER MANAGER
PRINTED CIRCUIT BOARD
1. Remove the Power Cartridge and unscrew the 4 screws (M3x12 Pan Head SEM Screw,
P/N 6974-312-SEQ) shown in Figure 56.
Figure 56: Battery Bridge Board Removal.
2. Remove the Unit Cover as described in the section Remove and Replace the Unit Cover.
CAUTION
The Eclipse contains electrostatic sensitive components. Do not open or
handle except at a static free workstation.
4
Screws