D
-
Eclipse S
™
Transportable Oxygen Concentrator with autoSAT
®
Eclipse 5 New Features
• NEW: All continuous flow and pulse flow settings available while operating on DC power
• NEW: Battery cartridge will re-charge when unit is connected to DC power for all continuous flow settings up to 2.0 lPM and all pulse flow settings 1 to 9
• NEW: Next generation internal components enhance unit reliability
Weight
Eclipse 5 Only: 15.0 lb (6.8 kg); Eclipse 5 w/ Battery Installed: 18.4 lb (8.4 kg)
Flow Rates
Continuous Flow
Pulse Flow
0.5 lPM
2.0 lPM
1.0 (16 ml)
3.0 (48 ml)
5.0 (80 ml)
8.0 (160 ml)
1.0 lPM
2.5 lPM
1.5 (24 ml)
3.5 (56 ml)
5.5 (88 ml)
9.0 (192 ml)
1.5 lPM
3.0 lPM
2.0 (32 ml)
4.0 (64 ml)
6.0 (96 ml)
2.5 (40 ml)
4.5 (72 ml)
7.0(128ml)
Battery Life
Continuous Flow
Pulse Flow
0.5 lPM -4.4 Hours 2.0 lPM -2.0 Hours 1.0 (16 ml) -5.4 Hours
4.0 (64 ml) -4.0 Hours
7 .0 (128 m L) -2.5 Hours
1.0 lPM -3.7 Hours 3.0 lPM -1.3 Hours 2.0 (32 ml) -5.1 Hours
5.0 (80 ml) -3.7 Hours
8.0 (160 m L) -2.0 Hours
3.0 (48 ml) -4.9 Hours
6.0 (96 ml) -3.5 Hours
9.0 (192 ml) -1.7 Hours
NOTE: Pulse flow times are calculated based on 12BPM. Higher breath rates will decrease battery times on pulse mode only.
Recharge Time
1. The typical time to recharge the battery inside the Eclipse on AC or
DC power is between 2 - 5 hours, dependent upon the flow setting.
The battery may be re-charged while the Eclipse is either ON or OFF.
2. The battery will recharge using the optional desktop charger in
approximately 3 hours.
3. It is acceptable to leave the Eclipse plugged into AC power at all
times while it is OFF. Note that while the unit remains plugged in
and OFF, the internal fan may continue to run while the battery is
charging and produce a faint humming noise.
Battery Gauge
The Eclipse 5 has a battery status gauge on the right side of the display
screen. Each of the 5 horizontal gray bars represents approximately 20%
of the total battery charge. When a battery is being charged, the indicator
bars will blink in a waterfall-type fashion. If a battery is not installed, or
is improperly installed, the battery status gauge will not appear on the
screen.
On/Off
Press and hold the ON/OFF button (top left Blue Button) for two seconds
to power on the Eclipse 5. Upon power up, the audible alarm should
sound for 3 seconds, and all three lights on above the display screen
should illuminate. It is normal for the yellow and/or red lights may remain
on for up to 5 minutes during warm-up. Once warm-up is complete, only
the green light should be on.
Press and hold the ON/OFF button for two seconds to power off the
Eclipse 5.
Changing Flow Mode
Press the flow mode button (bottom left blue button) to toggle between
pulse and continuous flow. The light adjacent to the button will illuminate
when in pulse mode and will blink with each breath detection and
delivery. If the unit is in pulse mode but does not detect a breath, it will
switch to continuous mode.