background image

A Chart Industries Company 

AirSep"  _/  SeQual" 

QUICK  REFERENCE 

Gu 

-

Eclipse S

Transportable Oxygen Concentrator with autoSAT

®

Eclipse 5 New Features 

• NEW: All continuous flow and pulse flow settings available while operating on DC power

• NEW: Battery cartridge will re-charge when unit is connected to DC power for all continuous flow settings up to 2.0 lPM and all pulse flow settings 1 to 9

• NEW: Next generation internal components enhance unit reliability

Weight 

Eclipse 5 Only: 15.0 lb (6.8 kg);  Eclipse 5 w/ Battery Installed: 18.4 lb (8.4 kg) 

Flow Rates 

Continuous Flow 

Pulse Flow 

0.5  lPM 

2.0  lPM 

1.0 (16 ml) 

3.0 (48 ml) 

5.0 (80 ml) 

8.0 (160 ml) 

1.0  lPM 

2.5  lPM 

1.5 (24 ml) 

3.5 (56 ml) 

5.5 (88 ml) 

9.0 (192 ml) 

1.5  lPM 

3.0  lPM 

2.0 (32 ml) 

4.0 (64 ml) 

6.0 (96 ml) 

2.5 (40 ml) 

4.5 (72 ml) 

7.0(128ml) 

Battery Life 

Continuous Flow 

Pulse Flow 

0.5  lPM -4.4 Hours  2.0  lPM -2.0 Hours  1.0 (16 ml) -5.4 Hours 

4.0 (64 ml) -4.0 Hours 

7 .0 (128 m L) -2.5 Hours 

1.0  lPM -3.7 Hours  3.0  lPM -1.3 Hours  2.0 (32 ml) -5.1 Hours 

5.0 (80 ml) -3.7 Hours 

8.0 (160 m L) -2.0 Hours 

3.0 (48 ml) -4.9 Hours 

6.0 (96 ml) -3.5 Hours 

9.0 (192 ml) -1.7 Hours 

NOTE: Pulse flow times are calculated based on 12BPM. Higher breath rates will decrease battery times on pulse mode only. 

Recharge Time 

1. The typical time to recharge the battery inside the Eclipse on AC or

DC power is between 2 - 5 hours, dependent upon the flow setting.

The battery may be re-charged while the Eclipse is either ON or OFF.

2. The battery will recharge using the optional desktop charger in

approximately 3 hours.

3.  It is acceptable to leave the Eclipse plugged into AC power at all

times while it is OFF. Note that while the unit remains plugged in

and OFF, the internal fan may continue to run while the battery is

charging and produce a faint humming noise.

Battery Gauge 

The Eclipse 5 has a battery status gauge on the right side of the display 

screen. Each of the 5 horizontal gray bars represents approximately 20% 

of the total battery charge. When a battery is being charged, the indicator 

bars will blink in a waterfall-type fashion. If a battery is not installed, or 

is improperly installed, the battery status gauge will not appear on the 

screen. 

On/Off 

Press and hold the ON/OFF button (top left Blue Button) for two seconds 

to power on the Eclipse 5. Upon power up, the audible alarm should 

sound for 3 seconds, and all three lights on above the display screen 

should illuminate. It is normal for the yellow and/or red lights may remain 

on for up to 5 minutes during warm-up. Once warm-up is complete, only 

the green light should be on. 

Press and hold the ON/OFF button for two seconds to power off the 

Eclipse 5. 

Changing Flow Mode 

Press the flow mode button (bottom left blue button) to toggle between 

pulse and continuous flow. The light adjacent to the button will illuminate 

when in pulse mode and will blink with each breath detection and 

delivery. If the unit is in pulse mode but does not detect a breath, it will 

switch to continuous mode. 

1-877-774-9271 | www.oxygenconcentratorstore.com

Reviews: