4. DISPLAY MODE AND OTHER INFORMATION MEASURING SHOWN ON THE SCREEN
Measurement display:
This display mode in addition to the "battery" icon is the default view of Cairsens.
It done a measure of the concentration of pollution in continue by the 3 left digits [1] and the 2 right
digits [2] exponent of the first 3 digits [1] is:
Value measured by the sensor = [XXX] * 10^ [xx].
For example:
Sensor measurement = 806 ppb
Display on screen = [806] [00] (ppb).
Sensor measurement = 8060 ppb
Display on screen = [806][01] (ppb).
The unit is ppb, and is represented by 3 characters [3] under 3 left digits [1].
The green LED [4] flashes regularly to indicate the proper functioning (continuously measures mode)
of Cairsens.
When a malfunction is detected by Cairsens, it will be displayed on the screen alternating with the
measurement every three seconds.
Display of malfunctions:
Detectable defects are:
cAL
(see photo cons): The sensor has reached its lifetime of 1 year. This timeout, measurement and calibration of
Cairsens are no longer certified. It is advisable to renew the Cairsens (Prices on request from your usual sales contact,
or email info@cairpol.com).
SEn
(see photo cons): The electrochemical cell is defective, contact the usual customer service or email
support@cairpol.com.
FAn
(see photo cons): The Cairsens fan is defective, contact the customer service department or by email to
support@ cairpol.com.
WARNING: This feature is under development and is not available in the current version!
5. CONFIGURATION CAIRSENS, DOWNLOADING, VIEWING AND EXPORT OF MEASURED DATA.
The configuration of certain parameters of Cairsens (Clock Set, name it, etc.) is possible using the CAIRFOT software (V4.5 and more).
CAIRSOFT download software
:
The Cairsoft is available on our web site '' www.cairpol.com '' under '' Download '' then '' Software '' or by clicking on this
link
.
The instruction manual is directly available from the icon
of Cairsoft home page or by clicking on this
Install the software “Cairsoft” on your computer (see the conditions and prerequisites in the instructions of Cairsoft).
Setup and configuration of Cairsens:
Connect the Cairsens to the PC using the specific USB to mini-USB cable provided (Do not use a commercially cable).
Follow the instructions of the User Guide of Cairsoft for its use.
The main features configurable in Cairsens are:
o
Date & Time (EU or US mode).
o
The name of Cairsens.
o
The choice of the measurement sampling rate (1 minute, 15 minutes or 1 hour) (note the induced shelf life).
Refer to Cairsoft User Guide for more details
View and export measurements to a PC:
The "graphic
” option of Cairsoft allows you to view and export in .XLSX file all data
stored by the Cairsens since its commissioning.
The choice of the viewing and export period is possible from a few days to several
months.
The data of Table “.XLSX” are the following:
[1] - {Cell B1} : Internal Reference of Cairsens
[2] - {Cell E1&F1}: Life sensor aging state in% translated as follows.
o
100% = 0 month of operation.
o
50% = 6 month of operation.
o
00% = 12 months of operation. In this case the corresponding
measures cells for in the table are red and the outline of the graphic
Cairsoft change from green to red (see the User Guide of Cairsoft).
{Column A2 to Ax}: Date and hour.
{Column B2 to Bx}: The pollutant measurements expressed in ppb (Or in µg/m
3
).
{Column C2 to Cx} : Battery voltage in mV translated as follows:
o
≥ 4700 mV = External power supply connected operation.
o
4200 <mV<3300 = Operation on battery without charging.
o
4200 mV = the battery is fully charged.
o
3300 mV = the battery is empty and must be recharged.
{Column D2 to Dx}: Temperature measurements inside the Cairsens (near the
electrochemical sensor) in ° C.
{Column E2 to Ex}: Relative humidity Measurements in the Cairsens (near the
electrochemical sensor) expressed in RH%.
[1]
[2]
[3]
Display example = 8060 ppb
Display example = 806 ppb
[2]
[1]
[3]
[4]
[4]
Tableau 1:
Export measures example
Cairsens O3/NO2
[1]
[2]