SERVICE MANUAL
CALIBRATION
3-15
Calculate the calibration constants:
VALUE1
= 240.0 /
MEAS1
VALUE2
= 240.0 /
MEAS2
VALUE3
= 240.0 /
MEAS3
VALUE4
= 240.0 /
MEAS4
VALUE5
= 240.0 /
MEAS5
Update the local high range frequency calibration of the TW by sending the following
GPIB string:
CAL
{1,2,3}
:OUTP:LOC:HIR:VOLT:FREQCAL
VALUE1 VALUE2 VALUE3 VALUE4 VALUE5
Close the TW’s output relay:
OUTPUT ON
Wait for the DMM reading to settle and record the value as
RDNG1
.
Query the TW’s output put voltage with the GPIB string below and record this value as
MEAS1
.
MEAS
{1,2,3}
:VOLT?
Set the TW’s frequency to 80 Hz with the following GPIB command:
SOUR:FREQ 80.0
Wait for the DMM reading to settle and record the value as
RDNG2
.
Query the TW’s output put voltage with the GPIB string below and record this value as
MEAS2
.
MEAS
{1,2,3}
:VOLT?
Set the TW’s frequency to 160 Hz with the following GPIB command:
SOUR:FREQ 160.0
Wait for the DMM reading to settle and record the value as
RDNG3
.
Query the TW’s output put voltage with the GPIB string below and record this value as
MEAS3
.
MEAS
{1,2,3}
:VOLT?
Set the TW’s frequency to 320 Hz with the following GPIB command:
SOUR:FREQ 320.0