SERVICE MANUAL
CALIBRATION
3-21
Set the TW’s frequency to 80 Hz with the following GPIB command:
SOUR:FREQ 80.0
Wait for the DMM reading to settle and record the value as
RDNG2
.
Query the TW’s output put voltage with the GPIB string below and record this value as
MEAS2
.
MEAS
{1,2,3}
:VOLT?
Set the TW’s frequency to 160 Hz with the following GPIB command:
SOUR:FREQ 160.0
Wait for the DMM reading to settle and record the value as
RDNG3
.
Query the TW’s output put voltage with the GPIB string below and record this value as
MEAS3
.
MEAS
{1,2,3}
:VOLT?
Set the TW’s frequency to 320 Hz with the following GPIB command:
SOUR:FREQ 320.0
Wait for the DMM reading to settle and record the value as
RDNG4
.
Query the TW’s output put voltage with the GPIB string below and record this value as
MEAS4
.
MEAS
{1,2,3}
:VOLT?
Set the TW’s frequency to 500 Hz with the following GPIB command:
SOUR:FREQ 500.0
Wait for the DMM reading to settle and record the value as
RDNG5
.
Query the TW’s output put voltage with the GPIB string below and record this value as
MEAS5
.
MEAS
{1,2,3}
:VOLT?
Set the TW’s frequency to 60 Hz and open the output with the following GPIB
commands:
SOUR:FREQ 60.0
OUTP OFF