CALIBRATION
TW SERIES
3-26
Wait for the counter timer to settle and record the phase angle as
RDNG1
.
Query the TW for the phase angle with the following command and record the value as
MEAS1
.
MEAS2:PHASE?
Set the TW frequency to 80 Hz via the GPIB:
SOUR:FREQ 80.0
Wait for the counter timer to settle and record the phase angle as
RDNG2
.
Query the TW for the phase angle with the following command and record the value as
MEAS2
.
MEAS2:PHASE?
Set the TW frequency to 160 Hz via the GPIB:
SOUR:FREQ 160.0
Wait for the counter timer to settle and record the phase angle as
RDNG3
.
Query the TW for the phase angle with the following command and record the value as
MEAS3
.
MEAS2:PHASE?
Set the TW frequency to 320 Hz via the GPIB:
SOUR:FREQ 320.0
Wait for the counter timer to settle and record the phase angle as
RDNG4
.
Query the TW for the phase angle with the following command and record the value as
MEAS4
.
MEAS2:PHASE?
Set the TW frequency to 500 Hz via the GPIB:
SOUR:FREQ 500.0
Wait for the counter timer to settle and record the phase angle as
RDNG5
.
Query the TW for the phase angle with the following command and record the value as
MEAS5
.
MEAS2:PHASE?