SERVICE MANUAL
CALIBRATION
3-39
3.28 HIGH RANGE CURRENT FREQUENCY CALIBRATION
Connect the 48 ohm resistor to the output of the TW through the power analyzer. Set
up the power analyzer to measure current.
Set up the TW to AC coupling, low range, remote sense, 40 Hz, 6.5 ampere current
limit, output relay closed and 240 volts with the following GPIB programming strings:
OUTP:COUP AC
SOUR:VOLT:RANGE HIGH
SOUR:SENSE REMOTE
SOUR:FREQ 40
SOUR
{1,2,3}
:CURR 6.5
OUTPUT ON
SOUR
{1,2,3}
:VOLT 240
Wait until the power analyzer has stabilized and record the current reading as
RDNG1
.
Query the TW’s current using the GPIB string below, and record this value as
MEAS1
.
MEAS
{1,2,3}
:CURR?
Program the TW to 80 Hz using the string below:
SOUR
{1,2,3}
:FREQ 80
Wait until the power analyzer has stabilized and record the current reading as
RDNG2
.
Query the TW’s current using the GPIB string below, and record this value as
MEAS2
.
MEAS
{1,2,3}
:CURR?
Program the TW to 160 Hz using the string below:
SOUR
{1,2,3}
:FREQ 160
Wait until the power analyzer has stabilized and record the current reading as
RDNG3
.
Query the TW’s current using the GPIB string below, and record this value as
MEAS3
.
MEAS
{1,2,3}
:CURR?
Program the TW to 320 Hz using the string below:
SOUR
{1,2,3}
:FREQ 320
Wait until the power analyzer has stabilized and record the current reading as
RDNG4
.
Summary of Contents for TrueWave TW1750
Page 2: ......
Page 4: ...ii This page intentionally left blank ...
Page 10: ...TABLE OF CONTENTS TW SERIES viii This page intentionally left blank ...
Page 36: ...MAINTENANCE AND TROUBLESHOOTING TW SERIES 2 16 This page intentionally left blank ...
Page 80: ...CALIBRATION TW SERIES 3 44 This page intentionally left blank ...