3-1
SECTION 3 – CALIBRATION
3.1 SCOPE
These procedures cover the calibration of TrueWave models TW1750, TW3500,
and TW5250. Calibrating the TrueWave requires no internal adjustment, therefore
the covers do not need to be removed (i.e., software calibration). Calibration is
accomplished by connecting test equipment to the unit’s output and entering externally
measured data via computer over the GPIB port.
3.2 APPLICABLE DOCUMENTS
TrueWave Operating Manual:
Elgar Document No. M161469-01
TrueWave SCPI Specification:
Elgar Document No. M161469-02
3.3 REQUIRED TEST EQUIPMENT
The equipment listed below or equipment with equivalent capabilities can be used to
calibrate the TrueWave's output and measurement. The measurement equipment listed
meets the 4:1 ratio required by many calibration standards.
Description
Accuracy
Model & Model No.
Digital Multimeter
6½ digits AC
AC Volts
Range: 0-350 VAC
Accuracy: 0.025%
Fluke 1271 with high
accuracy (option 12)
Power Analyzer,
three-phase
RMS Current
Range: 0-13 A/phase
Accuracy: 0.25%
Peak Current
Range: 0-56 Apk/phase
Accuracy: 1.25%
Power
Range: 0-1800W/phase
Accuracy: 0.625%
Voltech PM3000A
Counter Timer
Phase Angle
Accuracy: 0.25
°
Agilent 53131A
Resistor Divider
10:1 ration > 100K
Ω
Any
Resistive Loads
12
Ω
, 2000W (min.)
48Ω
, 2000W (min.)
Any
Any
Computer with
IEEE-488.2 Interface
N/A
Any
TABLE 3-1. CALIBRATION TEST EQUIPMENT
Summary of Contents for TrueWave TW5250
Page 2: ......
Page 4: ...ii This page intentionally left blank...
Page 10: ...TABLE OF CONTENTS TW SERIES viii This page intentionally left blank...
Page 36: ...MAINTENANCE AND TROUBLESHOOTING TW SERIES 2 16 This page intentionally left blank...
Page 80: ...CALIBRATION TW SERIES 3 44 This page intentionally left blank...