Should be snug enough so it does not slide along or
around the wrist.
•
The electrodes should be flush to the skin.
Electrode
Electrode
Right Handed
Device
Left Handed
Device
5.2 Calibrating Cala Trio
™
During stimulator setup, Cala Trio learns about
your tremor and personalizes therapy according
to its characteristics.
•
Calibration happens over the course of three
measurements taken while you perform the
"Tremor Task" prescribed by your physician.
Before calibrating, determine your "Tremor Task" by
checking your prescription information
Note:
In order to conserve battery, Cala Trio goes into
SLEEP
mode and fades to black if you are not actively
pressing any buttons. Press any button to wake
up Cala Trio.
Step 1:
Press the
MAIN
button to wake up the stimulator.
You should see Cala Trio displayed
UP
button
DOWN
button
MAIN
button
Step 2:
From Cala Trio, press and hold the
MAIN
button for
3 seconds to start setup. You will see "press main
to measure 1"
OR
Outstretched
postural hold
Wing-beating
postural hold
Step 3:
Get in position to do your "Tremor Task"
(
see above
)
You can find your "Tremor Task" in
your prescription information
Step 5:
Perform your “Tremor Task” until all dots are
blue
(~20 seconds)
. Cala Trio will vibrate when the
“Tremor Task” is complete
Step 4:
Press the
MAIN
button to start the measurement
Step 6:
If you correctly performed your "Tremor Task" for
the full measurement, press the
MAIN
button to
save the measurement
Note:
If you were moving in a way unrelated to your
tremor during the "Tremor Task", like walking or talking,
press the
DOWN
button and then the
MAIN
button to
remeasure.
Determine your "Tremor Task" by checking your
prescription information.