8
2. INSTALLATION AND MAPPING (to be carried out by the installer)
MONITOR 2.0 PULSE (pulse acquirer)
Firstly use the special wall mounting hooks provided, install the rear
plastic plate and, finally, use the fixing screws, as shown in the
following picture:
The generic pulse must be potential free
(volt free contact, maximum frequency 50 Hz.)
IN
GND
Fig. A
Pulse output
wire from the
volume meter
Connect the pulse output wire from the volume meter to the device
inputs with screw terminals (Figure A) as illustrated in the picture
below:
Positioning the consumption data acquisition device
In order to guarantee correct operation of the consumption data
acquisition device, it is necessary not to position the device in direct
contact with pipes nor under them, in order to avoid malfunctions due
to drops of condesation and humidity.
N.B.
When installation is complete,
the heat cost allocator is automatically activated after about 90 seconds
.
The following appears on the display:
You can now proceed with parameterisation of the heat cost allocator (see page 17).