Thermosight HT

temperature difference to the actual reading is indicated. That is why the indication an be positive or 
negative. 

7.6

HI (hi alarm)

The Hi alarm function is always active. Exceeding the adjusted alarm temperature (ex works setting 
1800°C) activates  an intermittent acoustic signal  (beep). Exceeding the end of the temperature 
range (> 1800°C) the frequency will be higher. 

7.7

CLR (clearing the stored measured values)

The clear function CLR deletes the complete data storage (data storage see 

5.2

). If the function is 

selected, one of the arrow keys  must be used to change the preselected value “no” into “yes”. 
Pushing the PAR button within the next 6 seconds will clear the storage. The display shows “run” 
until the storage is cleared.

7.8

STO

#

 (determine the storage position)

The STO function determines the storage position number (between 1 and 750) from which the py-
rometer starts to store measured values in the following data recording. The position number and 
the stored value of the last data logging will be displayed automatically. If a lower position number 
will be selected, all storage positions above this number will be overwritten by new values. If a high-
er position number will be selected, lower position numbers remain unused. Maximum 750 temper-
ature values can be stored, then the sign “STO FULL” will be blinking in the display.

7.9

RCL

#

 (recall measured values)

RCL shows the stored measured values. With the arrow keys the values can be displayed sequen-
tially.

7.10

INT (time interval for storage of measured values)

With the INT function it can be selected between single data storage or continuous data logging. In 
position single data storage (INT = OFF) only one temperature reading is stored by pushing the but-
ton into the second trigger point. In position continuous data logging (INT = 0.02 … 50) temperature 
readings are stored continuously as long as the pushbutton is in the second trigger position. The 
figures 0.02 … 50 are the time intervals in seconds between two stored values. Maximum 750 tem-
perature values can be stored, then the sign “STO FULL” will be blinking in the display.
Position INT = PRT (print) is used only in connection with the optional software 

PortaWin

. In this po-

sition a single temperature reading is stored on a connected PC by pressing the push button into 
the second trigger point, without using the built-in data logger of the pyrometer. 

7.11

C / F (temperature display in °C or °F)

The temperature can be displayed in °C or °F.

7.12

Time

The pyrometer is equipped with an integrated clock. Time and date are only used in combination 

- 11 -

Summary of Contents for Thermosight HT

Page 1: ...Thermosight HT Thermosight HT 1 User Manual...

Page 2: ...tions Setting keys 9 7 Parameter description adjustments 10 7 1 EMI emissivity 10 7 2 MAX maximum value 10 7 3 MIN minimum value 10 7 4 AVG average function 10 7 5 DIF difference temperature function...

Page 3: ...d in this manual or in case the content of this document should be incomplete or incorrect CALEX reserves the right to revise this document and to make changes from time to time in the content hereof...

Page 4: ...777 Temp exceeds 1800 C on display 8888 Display illumination automatically at dark ambient light conditions Measuring unit C F switchable Measuring functions Normal MAX MIN AVG DIF Alarm functions Hi...

Page 5: ...overview about all important safety aspects Additionally in the several sections there are concrete safety aspects to avert danger These as 5 Connector for RS232 interface and analog output Connector...

Page 6: ...sy alignment to the measuring object the pyrometers is equipped with a laser targeting light This is a visible red light with a wavelength between 630 and 680 nm and a maximum power of 1 mW The laser...

Page 7: ...can store up to 750 temperature values Should the pyrometer be switched to sin gle value data logging the data logging is confirmed by a beep and in the display the used number of the stored value is...

Page 8: ...he object has to be bigger or at least as big as the spot size of the pyrometer in the measuring distance For longer measuring distances than 800 mm the size of measuring spot can be calculated approx...

Page 9: ...ly displayed during adjustment Measuring functions are displayed additionally to the measuring temperature MAX MIN AVG DIF can give an acoustic signal HI or can be used for data analysis RCL Selected...

Page 10: ...um bright 0 1 Chromium bright 0 25 0 3 Brass oxidized 0 6 0 7 Bronze bright 0 03 Copper oxidized 0 7 0 85 Measuring object EMI 1 45 1 8 m Zinc 0 45 0 55 Nickel 0 15 0 2 Gold Silver bright 0 02 Porcela...

Page 11: ...ber will be selected lower position numbers remain unused Maximum 750 temper ature values can be stored then the sign STO FULL will be blinking in the display 7 9 RCL recall measured values RCL shows...

Page 12: ...r 1 mV F and can be used e g for connection to a line recorder Via serial interface the pyrometer can be connected to a PC This allows to read out the stored data and analyze them graphically or in ta...

Page 13: ...ys of the pyrometer Display Description Battery capacity low only short operation time of the pyrometer automatically shut down on too low capacity 7777 Measuring temperature below beginning of temper...

Page 14: ...o achieve the max temperature signal Measuring object is smaller than spot size Choose correct measuring distance Measuring object is not always in the measuring spot of the pyrometer e g swinging wir...

Page 15: ...osight HT Accessories AS 101 Close up lens AS 102 Software inkl RS 232 cable AS 103 Analogue cable AS 104 Charger for 9V Accu NiMH AS 106 NiMH Accu 9V AS 107 Tripod AS 109 Belt Pack AS 111 Carrying Ca...

Page 16: ...Locking lever 7 M Maintenance 14 Maximum value 10 Minimum value 10 P Push button 7 S Scope of delivery 5 Sighting 7 Software PortaWin 12 Special displays 13 Spot size 7 Storage 11 T Time 11 Trouble sh...

Reviews: