background image

Thermosight HT

with the stored values and can only be read out via interface and software  

PortaWin

. On the PC 

they are shown additionally to the measured temperatures (see 8.2 software 

PortaWin

).

The clock has its own power supply and is running even if the pyrometer battery is taken out. 
Time setting: - pyrometer must be switched off

- press laser targeting light key and parameter key at the same time
- keep them hold and press the push button to switch the instrument on.
- time appears on the display
- the blinking part, hours or minutes can be set with the arrow keys 
- switch from minutes to hours and back with the parameter key
- switch back to the measuring mode by pushing the trigger switch

The date can only be changed by PC via software 

PortaWin

, not on the instrument.

8

Interface

The pyrometer is equipped with an analog output and 
a serial RS232 interface (baud rate 57600 Bd).
 
The analog output is 1 mV / °C (or 1 mV / °F) and can 
be used e.g. for connection to a line recorder.
Via serial interface the pyrometer can be connected 
to a PC. This allows to read out the stored data and 
analyze them graphically or in table form. Also the in-
strument can be used in continuous mode in combi-
nation with a PC and the data are displayed on the 
monitor as an online diagram (for connecting cables 
and   software   see  

13   Reference   numbers   acces-

sories

).

8.1

Connector pin assignment

For use of analog or digital interface a suitable connecting cable is necessary (reference numbers 
see  

13 Reference numbers accessories

).

Interface:

Analog output:

 

PC side 

(soldering side)

 

1

 

2

 

3

 

4

 

5

 

6

 

7

 

8

 

9

 

Pyrometer’s side

 

cable connection

 

1

 

2

 

3

 

4

 

 

Pyrometer’s side 

1

 

2

 

3

 

4

 

–U

A

 

+U

A

 

8.2

Software 

PortaWin

 (accessory)

PortaWin

 is a Windows based software which enlarges the functions of the pyrometer, the min. re-

quirement is Windows 95. To install the software, double-click the setup.exe file on the installation 
CD. Starting the program will show the following icons with the corresponding functions:

- 12 -

connector socket 
USB, type B 
for analog output 
and interface

Summary of Contents for Thermosight HT

Page 1: ...Thermosight HT Thermosight HT 1 User Manual...

Page 2: ...tions Setting keys 9 7 Parameter description adjustments 10 7 1 EMI emissivity 10 7 2 MAX maximum value 10 7 3 MIN minimum value 10 7 4 AVG average function 10 7 5 DIF difference temperature function...

Page 3: ...d in this manual or in case the content of this document should be incomplete or incorrect CALEX reserves the right to revise this document and to make changes from time to time in the content hereof...

Page 4: ...777 Temp exceeds 1800 C on display 8888 Display illumination automatically at dark ambient light conditions Measuring unit C F switchable Measuring functions Normal MAX MIN AVG DIF Alarm functions Hi...

Page 5: ...overview about all important safety aspects Additionally in the several sections there are concrete safety aspects to avert danger These as 5 Connector for RS232 interface and analog output Connector...

Page 6: ...sy alignment to the measuring object the pyrometers is equipped with a laser targeting light This is a visible red light with a wavelength between 630 and 680 nm and a maximum power of 1 mW The laser...

Page 7: ...can store up to 750 temperature values Should the pyrometer be switched to sin gle value data logging the data logging is confirmed by a beep and in the display the used number of the stored value is...

Page 8: ...he object has to be bigger or at least as big as the spot size of the pyrometer in the measuring distance For longer measuring distances than 800 mm the size of measuring spot can be calculated approx...

Page 9: ...ly displayed during adjustment Measuring functions are displayed additionally to the measuring temperature MAX MIN AVG DIF can give an acoustic signal HI or can be used for data analysis RCL Selected...

Page 10: ...um bright 0 1 Chromium bright 0 25 0 3 Brass oxidized 0 6 0 7 Bronze bright 0 03 Copper oxidized 0 7 0 85 Measuring object EMI 1 45 1 8 m Zinc 0 45 0 55 Nickel 0 15 0 2 Gold Silver bright 0 02 Porcela...

Page 11: ...ber will be selected lower position numbers remain unused Maximum 750 temper ature values can be stored then the sign STO FULL will be blinking in the display 7 9 RCL recall measured values RCL shows...

Page 12: ...r 1 mV F and can be used e g for connection to a line recorder Via serial interface the pyrometer can be connected to a PC This allows to read out the stored data and analyze them graphically or in ta...

Page 13: ...ys of the pyrometer Display Description Battery capacity low only short operation time of the pyrometer automatically shut down on too low capacity 7777 Measuring temperature below beginning of temper...