User Manual
2253i / 2253iX
AMETEK
California Instruments
153
Number Message String
Cause
Remedy
-314
"Save/recall memory
lost"
User setup register
contents lost
Store setup in same register again.
-315
"Configuration memory
lost"
Hardware configuration
settings lost.
Contact CI service department at
support@programmablepower.com
to
obtain instructions on restoring
configuration data.
-330
"Self-test failed"
Internal error
Contact CI service department at
support@programmablepower.com
-350
"Queue overflow"
Message queue full.
Too many messages. Read status using
SYST:ERR query until 0, "No Error" is
received indicating queue empty.
-400
"Query error"
Unable to complete query.
Check programming manual for correct
query format and parameters
-410
"Query INTERRUPTED"
Query issued but
response not read.
Check application program for correct
flow. Response must be read after each
query to avoid this error.
-420
"Query
UNTERMINATED"
Query incomplete.
Check for terminator after query
command.
-430
"Query DEADLOCKED"
Query cannot be
completed
Check application program for multiple
queries
-440
"Query
UNTERMINATED"
Query incomplete.
Check for terminator after query
command.
0
"No error"
No errors in queue
40
"Voltage self test error,
output 1”
No. 1/A amplifier in
Master source has no
output during Self-test.
Contact CI service department at
support@programmablepower.com
41
"Voltage self test error,
output 2”
No. 2/B amplifier in
Master source has no
output during Self-test.
Three phase models only.
Contact CI service department at
support@programmablepower.com
42
"Voltage self test error,
output 3”
No. 3/C amplifier in
Master source has no
output during Self-test
Three phase models only.
Contact CI service department at
support@programmablepower.com
43
"Current self test error,
output 1”
No. 1/A amplifier in Aux.
Source has no output
during Self-test.
Contact CI service department at
support@programmablepower.com
44
"Current self test error,
output 2”
No. 2/B amplifier in Aux.
Source has no output
during Self-test. Three
phase models only.
Contact CI service department at
support@programmablepower.com
45
"Current self test error,
output 3”
No. 3/C amplifier in Aux.
Source has no output
during Self-test. Three
phase models only.
Contact CI service department at
support@programmablepower.com
403
"CAL not enabled"
No password entered for
calibration
Enter correct CAL password.
801
"Output volt fault"
- Output voltage does not
match program value
when ALC is on.
- Over load
- Voltage kick-back
- No output voltage
Load exceeds current limit and unit is in
Constant Voltage (CV) mode of
operation.
- Reduce load or increase CL setting.
Output voltage is driven above
programmed voltage by external