INSTRUCTIONAL MANUAL
2750L-P/4500L-P/6000L-P
August, 2001
Rev. K
-34-
The following are for measured values.
NO SCREEN
NAME
EXTENSIONS
ARGUMENT
ACTION TAKEN
11
ELT
H, M, S
Hrs,Min,Sec
Reports the total accumulated run time up to
9,999 hours.
21
VLT
A, B, C
0-400.0
Measures the TRMS output voltage.
22
CUR
A, B, C
0.0-200.0
or
0.00-20.00
Measures the TRMS output current in Amps.
Range depends on maximum current per
phase.
23
PWR
A, B, C
0.00-27.00KW or
0-2000 watts
Measures the True output power. Range
depends on output VA per phase
24
PWF
A, B, C
0-1.000
Measures the power factor of the load.
25
APW
A, B, C
0.00-27.00KVA or
0-2000 VA
Measures the apparent out put power. Range
depends on output VA per phase.
26
FQM
40-5000
Measures the output frequency in hertz
27
PZM
A, B, C
0-359.9
Measures the phase angle of the output voltage
relative to phase A and phase A relative to an
external input
Table 3-6
MEASUREMENT SCREENS
NO SCREEN
NAME
EXTENSIONS
ARGUMENT
ACTION TAKEN
13
CAL VLT
A, B, C
Actual out put
voltage
Calibrates the measured voltage to be the
same as argument
14
CAL CUR
A, B, C
Actual out put
current (amps)
Calibrates the measured current to be same as
argument
15
CAL PWR
A, B, C
Actual out put
power
Calibrates the measured power to be same as
argument. The argument is in KW for power
systems of more than 2000 VA per phase
20
POF
A, B, C
0-±359.9
Calibrates the programmed output phase a
Table 3-7
CALIBRATION SCREEN