OPTIONS SPECIFICATIONS
Universal Input Power (UP option):
342 to 456 V
L-L
three phase
Output Voltage Ranges:
156/312 (HV option)
67.5/135
(LV
option)
200.0/400.0
(EHV
option)
Master:
Must be used with LK option
Slave:
Must be used with LK option
Square Wave (4500L-3P and 2750L-3P only) Performance from 50 Hz to 5 KHz:
Rise and Fall Time (10% to 90%):
Less than 40 microseconds
Droop:
Less than 10%
Overshoot (resistive load):
Less than 20%
Mode: Allows power system to be programmed to a 1 or 3 phase output.
Auxiliary outputs (Option AX):
Master and Slave options not available with this option
(All specifications at 23
°
C and verified with TRMS voltmeter. The Phase D sense lines must be connected to
the Phase D output.)
Phase
D:
Voltage:
26.0
Vrms
fixed
Accuracy:
±2%
Current:
3.0
Amps
(Max)
Frequency:
360
to
440
(Tracks the main outputs)
Phase:
Tracks
the
Phase
A
output
Accuracy:
±1 degree. (All outputs at half load, 400 Hz)
Distortion:
1% THD maximum with linear load
Line
Regulation:
±0.05%
using
external
sense
Load Regulation:
±0.05% using external sense
Phase
E
Voltage:
5.0
Vrms
fixed
Accuracy:
±5%
Current:
5.0
Amps
(Max)