ADC 2
Instruction Manual, November 2016
Page 8
Further requirements:
•
Do not place the instrument in a location where the tempera-
ture significantly changes (e.g. under an air conditioning duct
or by a window). Significant changes in temperature will affect
the performance of the unit
•
Do not place the instrument in direct sunlight. Direct sunlight
will disturb the measurement of the CCD sensor
•
Do not use the instrument in an environment with moving am-
bient air (draft)
•
Do not expose the instrument to any strong vibration or shock
•
Avoid placing the instrument near equipment that radiates
heat. Do not place the instrument near gas burners, electric
heaters or ovens
•
Do not place the instrument near equipment that generates in-
tense magnetic fields such as electric welding equipment, high
frequency furnaces, pole transformers, etc.
•
Protect the instrument from excessive dust
•
Connect the instrument to power lines that are free from sud-
den changes or voltage fluctuations
•
If you must use power motor driven equipment (such as a stir-
rer or shaker) in the same line as your instrument, ensure that a
noise reduction unit is in the same power line
2.3.
Conditions for the installation
Confirm that the following requirements exist before installing the
instrument:
Power supply and ground
Line voltage:
100 – 240V (see rating plate on instrument).
Frequency:
50 / 60 Hz
Power capacity:
20 W
Ground terminal:
A grounded outlet should be located within 2 me-
ters of the instrument.
Connection for Exhaust tube