Introduction 

Instruction manual ATS4, Mar-16 

3

 

1

 

Introduction 

1.1

 

CAMAG automatic TLC sampler 4 (ATS4) 

The CAMAG Automatic TLC Sampler 4 is a fully automatic sample application device for 
use in qualitative, quantitative and preparative Thin-layer chromatography. The system 
consists of the instrument with interface, computer and software from which the 
instrument is programmed. 
The parameters for up to 6 application programs can be saved for stand-alone operation 
without PC- control. 
The ATS4 is controlled by CAMAG software that manages, monitors and reports all 
parameters and steps of the TLC analysis including definition of plate material, sample 
application, derivatization, development and evaluation. All data is handled in a GMP/GLP-
compliant manner. 

1.2

 

Precautions 

 

Please read this operating manual before starting the installation!

 

This manual contains information and warnings the user has to follow to ensure 
reliable operation of the instrument.  

 

Some interior parts of the instrument are under AC power. Careless 

and improper use can cause injury. Unauthorized manipulations can 
cause damage 

 

This sign indicates (on instrument and in this manual) that failure to take note of 
the accompanying information may result in damage of the instrument 

 

The instrument is manufactured and tested in accordance with the respective 
European safety publications shown on the Declaration of Conformity (DoC). The 
instrument complies with safety class 1 and has been designed for indoor use only 
(IP 20). Further, this device has passed the CAMAG Quality Assurance tests and has 
been delivered in safe operation condition. For detailed instrument data see chapter 
technical data 

 

Attention: For safety reasons the instrument may only be used for the purposes 
described in the operating manual 

 

To avoid injury use adequate safety equipment (protective goggles, gloves etc. if 
applicable) when working with the instrument  

 

For the sample rack 022.7430 only standard 2 mL 12x32 mm with snap-cap, i.e. 
crimp closure or compatible are to be used. The penetration depth of the needle is 
fixed. Vials with the wrong depth may cause damage to the needle 

Summary of Contents for ATS 4

Page 1: ...INSTRUCTION MANUAL AUTOMATIC TLC SAMPLER 4...

Page 2: ...n 9 Rinsing bottles 9 Gas air connection 10 Dosage syringe and needle 10 Mains connection 11 Connection to the PC and installation 12 The user dialog 12 Set baud rate 12 Set display language 13 Set LC...

Page 3: ...Mounting the baffle bridge with septum foil 19 5 Maintenance and troubleshooting 21 5 1 Handling of the HAMILTON syringe 21 5 2 Adjusting the spray gas pressure 21 5 3 Cleaning the spray nozzle 21 5 4...

Page 4: ...nterior parts of the instrument are under AC power Careless and improper use can cause injury Unauthorized manipulations can cause damage This sign indicates on instrument and in this manual that fail...

Page 5: ...efore installation and use the instrument should be acclimated properly Use a damp lint free cloth for cleaning the instrument surface Do not employ aggressive detergents Protect yourself and the inst...

Page 6: ...lectrical and electronic equipment WEEE Directive 2002 96 EC To properly recycle the instrument or parts of it you are requested to send the equipment back to the distributor producer or an adequate c...

Page 7: ...yringe needle for contact application red 695 0065 Replacement seal for syringe needle 695 0046 47 Pkg of 5 695 0061 Replacement piston for 10 L syringe 695 0062 Replacement piston for 25 L syringe 69...

Page 8: ...and lift the instrument up carefully Never try to pull or lift the instrument by means of the transparent cover The cover is not connected with the instrument and will break Do not switch on power bef...

Page 9: ...e from acidic alkaline or other gas that may corrode metal or painted surfaces must be secured Other requirements Do not place the instrument in a location where the temperature undergoes significant...

Page 10: ...uirements exist before installing the instrument Power supply and ground Line voltage 85 250VAC see rating plate on instrument Frequency 47 63 Hz Power capacity 60 VA A grounded outlet should be withi...

Page 11: ...pressure to 6 8 bar 87 115 psi The right air connection is the outlet of the vacuum pump and does not need to be connected Dosage syringe and needle Determine the proper syringe needle for the specifi...

Page 12: ...e is correctly in place After mounting the syringe the syringe volume and the needle type must be entered into the instrument Procedure Press the DIALOG key on the instrument and choose with the keys...

Page 13: ...dialog you can select the communication Baud rate between the instrument and the PC the LCD language and contrast To enter the user dialog Hold down the DIALOG key and press the RESET key Release the...

Page 14: ...e setting Press the ENTER key to accept the current setting and move on to the LCD contrast selection Press the RESET key to quit the user dialog Set LCD contrast Start the user dialog according to th...

Page 15: ...objects in the instrument for instance below the sample rack 3 2 Check syringe and needle type Check if the syringe size and needle type is appropriate for your analysis method If not change the syri...

Page 16: ...t it touches the rear stage ledge and the positioning pin 4 in the middle of the stage Flip up the front stage ledge 2 and move the plate lift lever 1 back to the front Guide the lever with your hand...

Page 17: ...o below figure and may be left in place if only foils are being used The positioning of foils is similar to that of TLC plates Note also that the foil must be placed underneath the down holder and fir...

Page 18: ...rrupts the current application Starts the manual operations dialog to rinse the syringe reset the waste position or clean the rinsing container and spray nozzle RESET Resets and re initializes the ins...

Page 19: ...ection with ENTER The syringe is now filled with rinsing solvent 25 of this solvent is dosed into the spray nozzle and allowed to pour down into the waste container Then the spray nozzle is blown clea...

Page 20: ...d the type of the needle Accept your selection with ENTER 4 9 Stopping or resetting the instrument Press the RESET key This will abort all activities in progress and re initialize the instrument 4 10M...

Page 21: ...septum holder 5 attachment screws Position the new baffle bridge with septum foil and attach it with the two screws 5 Insert a septum foil 3 with the Teflon layer face down into the groove of the baff...

Page 22: ...bodies are exchangeable CAMAG recommends exchanging those parts regularly Check the CAMAG Maintenance Data Sheet for information of replacement cycles 5 2 Adjusting the spray gas pressure Spray gas f...

Page 23: ...ply samples from open vials use the optional baffle bridge with septum foil CAMAG part number 022 7460 see Fig 9 The septum foil CAMAG part number 022 7462 is inserted with the Teflon layer face down...

Page 24: ...wer cord Fig 10 Fuse holders 5 7 Troubleshooting During the filling of the syringe bubbles are visible The connection between syringe and needle is not tight Action check connection and tighten it The...

Page 25: ...ght Action Check lid and tighten it No syringe present Action Insert syringe and attach tubing Vacuum pump or valve for emptying waste container is defective Action Request service for vacuum pump or...

Page 26: ...ontact 220 Rack drive Movement over position limit was requested and aborted by low level driver Call your service contact 221 Stage drive Movement over position limit was requested and aborted by low...

Page 27: ...tto Call your service contact 1147 Lift drive ditto Call your service contact 1148 Syringe drive ditto Call your service contact 1154 Rack drive No start Call your service contact 1155 Stage drive dit...

Page 28: ...mm steps X drive stage Step motor 3200 steps rev 8 steps 0 1 mm Maximum speed 200 mm s with acceleration ramp Positions programmable 5 0 195 0 mm in 0 1 mm steps Gas supply External pressure 4 5 6 bar...

Page 29: ...sumable parts Part No Description Replacement cycle Heavy frequent occasional use 695 0061 Plunger 10 l syringe 3 6 12 Months 695 0062 Plunger 25 l syringe 3 6 12 Months 695 0063 Plunger 100 l syringe...

Page 30: ...61 467 34 34 Fax 41 61 461 07 02 E Mail info camag com CAMAG Germany Bismarckstra e 27 29 D 12169 Berlin Telephone 49 30 516 55 50 Fax 49 30 795 70 73 E Mail infoberlin camag com CAMAG Scientific Inc...

Page 31: ......

Page 32: ..._____________________________________________________________ __________________________________________________________________________________ _______________________________________________________...

Page 33: ..._____________________________________________________________ __________________________________________________________________________________ _______________________________________________________...

Reviews: