Introduction
Instruction manual ATS4, Mar-16
4
•
Before first operation, check whether the voltage shown on the instrument
matches your local mains voltage. The power cord may only be connected to a
grounded, fused (not higher than 16A) outlet. Do not use extension cords without
ground contact
•
The exhaust of the internal vacuum pump is located at the rear of the instrument. It
can be connected to a fume hood
•
When working with the fluids of the instrument, be sure to take the appropriate
caution (protect your eyes from direct contact with liquid)
•
In order to avoid damage to the bottle and syringe connections these have to be
tightened by hand only
•
Avoid overflow of the waste bottle. The vacuum pump is damaged and liquid can
spill
•
Risk of finger squeeze between moving parts; be careful closing the door
•
The instrument may be used only by properly trained laboratory staff
•
The instrument may not be used in rooms with danger of explosions
•
The instrument contains highly sophisticated electronics and optical parts. It may be
operated only in a non-condensing atmosphere in the temperature range outlined
in the chapter "Technical Data". Before installation and use, the instrument should
be acclimated properly
•
Use a damp lint free cloth for cleaning the instrument surface. Do not employ
aggressive detergents
•
Protect yourself and the instrument from electrostatic shock which can cause
damage to the electronic parts
•
Only authorized personnel may open the instrument. Service and repair is only to
be performed by trained specialists. Use spare parts and consumables supplied by
CAMAG only. The warranty is voided if parts from other sources are used. Check
the service manual before you start service to reduce product-specific risks
•
The power cord has to be removed before the instrument is opened. It is not
permitted to work on an instrument that has been opened and is connected to the
power supply
•
Spare fuses must be of the type specified by the instrument manufacturer. It is
forbidden to short-circuit or manipulate fuses
•
If the instrument is found to be defective, it must be switched off and steps must
be taken to ensure that it cannot be switched on by mistake
•
If liquids penetrate the inside of the instrument, the power has to be disconnected
immediately. Small amounts of liquid can be wiped off and/or dried by means of a
hairdryer, with larger amounts of liquid a service technician has to be called. A test
of functionality has to be performed in all cases