Manual operation
Instruction manual ATS4, Mar-16
18
•
Press the RUN key
•
The most recently used method is displayed. With the
and
keys you can select
another locally saved method.
•
Press the ENTER or RUN key to start the application.
If no method is locally saved the message “MEMORY EMPTY” is displayed for 2 second.
Then the instrument goes to ready position.
4.3
Aborting an application in progress
•
Press the END key
The application in progress is aborted. An aborted application cannot be completed later.
4.4
Rinsing the syringe and spray nozzle
Syringe and spray nozzle can only be rinsed if no application is in progress and a syringe is
installed.
•
Press the END key
•
Use the
and
keys to select the function „RINSE SYRINGE“
•
Accept the selection with ENTER.
The syringe is now filled with rinsing solvent. 25% of this solvent is dosed into the spray
nozzle and allowed to pour down into the waste container. Then the spray nozzle is
blown clean with N2. This procedure is repeated 3 times.
4.5
Reset the pre-dosage position
The pre-dosage position can only be reset if no application is in progress.
•
Press the END key
•
Use the
and
keys to select the function „NEW WASTE PLATE“
•
Accept the selection with ENTER.
The pre-dosage position is now reset to the first position on a new 20x10cm TLC plate.
Always use a 20x10cm TLC plate as waste plate.
4.6
Clean the rinsing unit
The rinsing unit can only be cleaned if no application is in progress and a syringe is
installed.
•
Press the END key
•
Use the
and
keys to select the function „CLEAN RINSE UNIT“
•
Accept the selection with ENTER.
This way the rinse unit will be emptied through the syringe and then refilled with fresh
rinsing liquid.