Technical data
Instruction manual ATS4, Mar-16
27
6
Technical data
Power connections
85 – 250 V
∼
47 – 63 Hz 60 VA
Dosing syringe
Choice of 10, 25 or 100 µL, gas-tight with side port
Dosing volume
100 nL to 1 mL in steps of 100nL
Syringe drive
Stepper motor 1600 steps / rev.:
100nL = 960 steps for 10µL syringe
100nL = 384 steps for 25µL syringe
100nL = 96 steps for 100µL syringe
Y-drive
(turret)
Stepper motor 1600 steps/rev. 4 steps = 0.1 mm
Maximum speed 200 mm/s with acceleration ramp
Positions programmable: 5.0 – 195.0 mm in 0.1 mm steps
X-drive
(stage)
Step motor 3200 steps/rev. 8 steps = 0.1 mm
Maximum speed 200 mm/s with acceleration ramp
Positions programmable: 5.0 – 195.0 mm in 0.1 mm steps
Gas supply
External pressure: 4.5 - 6 bar (65 - 87 psi).
Consumption:
0.2 - 0.3 L/min for contact application
0.8 – 2.0 L/min for spray-on application (pressure dependent)
Pressure rinsing bottle
100 mbar above normal pressure
Pressure waste bottle
100 mbar below surrounding pressure
Septum puncher
Suitable for normal septa; pneumatic operation
Reservoir for rinsing liquid
250 mL bottle
Rack
For 66 sample vials 2 mL (12x32mm) or 96-well-plate (optional)
Max. no of application
100 application per analysis
Measures
Width 630mm, depth 530mm, height 500mm
Weight
36 kg